CardX - Cardiac Implantable Electronic Devices, published by HL7 International / Clinical Interoperability Council. This guide is not an authorized publication; it is the continuous build for version 1.0.0 built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/CardX-CIED/ and changes regularly. See the Directory of published versions

Resource Profile: Observation - CIED Connectivity

Official URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-connectivity-status Version: 1.0.0
Standards status: Trial-use Active as of 2025-12-15 Maturity Level: 2 Computable Name: CIEDConnectivityStatus

Profile of the observation resource to present the most recent connectivity status and associated information for a CIED/Monitor combination. In most instances there will be one observation per CIED/monitor pair per patient, however some manufacturers may allow multiple CIEDs implanted in a patient to connect to a single monitor and have a single connectivity status for the entire system of devices.

CIEDs and monitors connect at regular intervals to exchange data ranging from every few minutes to every other week depending on the device/monitor configuration and manufacturer. The connectivity observation profile is a snapshot in time and is not intended to be used to query historical log information that may have been captured by the remote monitoring IoT platform. A single observation will be returned for every CIED/monitor pair.

Connectivity status is based on the actual time difference between the most recent successful communication and the present date. To manage disconnected patients, manufacturers have implemented additional functionality in their platforms. The connectivity modifier is included to represent an action performed by the user or business logic applied by the manufacturer. Examples of these are a user suspending a patient from appearing on the disconnected list in the remote monitoring system or the system automatically marking a patient as inactive after a prolonged disconnection period to remove them from appearing on the disconnected list in the user interface.

Usages:

You can also check for usages in the FHIR IG Statistics

Formal Views of Profile Content

Description Differentials, Snapshots, and other representations.

NameFlagsCard.TypeDescription & Constraints    Filter: Filtersdoco
.. Observation C 0..* Observation Measurements and simple assertions
Constraints: obs-6, obs-7, obs-8
... implicitRules ?!Σ 0..1 uri A set of rules under which this content was created
... contained 0..* Resource Contained, inline Resources
... Slices for extension 1..* Extension Extension
Slice: Unordered, Open by value:url
.... extension:ciedDevice S 1..* Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) Cardiac implantable device(s)
URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension
... status ?!SΣ 1..1 code registered | preliminary | final | amended +
Binding: ObservationStatus (required): Codes providing the status of an observation.
... code SΣC 1..1 CodeableConcept Type of observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
.... id 0..1 string Unique id for inter-element referencing
.... extension 0..* Extension Additional content defined by implementations
.... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
..... version 0..1 string Version of the system - if relevant
..... code 1..1 code Symbol in syntax defined by the system
Fixed Value: observation-cied-connectivity
..... display 0..1 string Representation defined by the system
..... userSelected 0..1 boolean If this coding was chosen directly by the user
.... text 0..1 string Plain text representation of the concept
... subject SΣ 1..1 Reference(Patient - CIED Patient Profile) Who and/or what the observation is about
... effective[x] SΣ 1..1 dateTime Clinically relevant time/time-period for observation
... value[x] ΣC 0..1 Actual result
.... valueQuantity Quantity
.... valueCodeableConcept CodeableConcept
.... valueString string
.... valueBoolean boolean
.... valueInteger integer
.... valueRange Range
.... valueRatio Ratio
.... valueSampledData SampledData
.... valueTime time
.... valueDateTime dateTime
.... valuePeriod Period
.... valueAttachment Attachment
.... valueReference Reference(MolecularSequence)
... device S 0..1 Reference(Device - CIED Monitor Profile) A reference to the device that generates the measurements or the device settings for the device
... Slices for component SΣC 1..* BackboneElement CIED Connectivity Components
Slice: Unordered, Open by value:code
.... component:All Slices Content/Rules for all slices
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
..... value[x] SΣ 0..1 Actual component result
...... valueQuantity Quantity
...... valueCodeableConcept CodeableConcept
...... valueString string
...... valueBoolean boolean
...... valueInteger integer
...... valueRange Range
...... valueRatio Ratio
...... valueSampledData SampledData
...... valueTime time
...... valueDateTime dateTime
...... valuePeriod Period
...... valueAttachment Attachment
...... valueReference Reference(MolecularSequence)
.... component:connectivityStatus SΣC 1..1 BackboneElement The overall system connectivity status as calculated by the CIED manufacturer.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-status
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Value Set (required)
.... component:connectivityModifier SΣC 0..* BackboneElement Additional modifiers to connectivity status as calculated by the CIED manufacturer.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-modifier
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Modifier Value Set (required)
.... component:statusReason SΣC 0..1 BackboneElement Text describing reasons why the patient may be listed at a given status and/or modifier.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-reason
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 string Actual component result
.... component:statusGuidance SΣC 0..1 BackboneElement Text describing steps to get the patient to a connected status
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-guidance
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 string Actual component result
.... component:lastCiedConnectivityDate SΣC 0..1 BackboneElement The timestamp of the most recent communication between the CIED and monitor.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-cied-connectivity-time
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:lastMonitorConnectivityDate SΣC 0..1 BackboneElement The timestamp of the most recent communication from the monitor to the remote monitoring platform.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-monitor-connectivity-time
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:nextCiedConnectivityDate SΣC 0..1 BackboneElement The timestamp of the next expected communication between the CIED and monitor.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-cied-connectivity-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:nextMonitorConnectivityDate SΣC 0..1 BackboneElement The timestamp of the next expected communication from the monitor to the remote monitoring platform.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-monitor-connectivity-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:lastRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the most recent remote device interrogation with a personal monitor.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:nextScheduledRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the next scheduled remote device interrogation.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:nextSecondaryScheduledRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-secondary-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:connectionInterval SΣC 0..1 BackboneElement The time interval in days between expected CIED and remote monitoring software communication.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connection-interval
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 integer Actual component result
.... component:expectedTimePeriod SΣC 0..1 BackboneElement The time period in days in which successful communication must take place for the connectivity status to remain connected.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: expected-time-period
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 integer Actual component result

doco Documentation for this format

Terminology Bindings

Path Status Usage ValueSet Version Source
Observation.status Base required Observation Status 📍5.0.0 FHIR Std.
Observation.code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​value[x] Base required CIED Connectivity Status Value Set 📦1.0.0 This IG
Observation.component:connectivityModifier.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityModifier.​value[x] Base required CIED Connectivity Status Modifier Value Set 📦1.0.0 This IG
Observation.component:statusReason.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:statusGuidance.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastCiedConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastMonitorConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextCiedConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextMonitorConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextScheduledRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectionInterval.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:expectedTimePeriod.​code Base example LOINC Codes 📍5.0.0 FHIR Std.

Constraints

Id Grade Path(s) Description Expression
dom-2 error Observation If the resource is contained in another resource, it SHALL NOT contain nested Resources contained.contained.empty()
dom-3 error Observation If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().ofType(canonical) | %resource.descendants().ofType(uri) | %resource.descendants().ofType(url))) or descendants().where(reference = '#').exists() or descendants().where(ofType(canonical) = '#').exists() or descendants().where(ofType(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
dom-4 error Observation If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
dom-5 error Observation If a resource is contained in another resource, it SHALL NOT have a security label contained.meta.security.empty()
dom-6 best practice Observation A resource should have narrative for robust management text.`div`.exists()
ele-1 error **ALL** elements All FHIR elements must have a @value or children hasValue() or (children().count() > id.count())
ext-1 error **ALL** extensions Must have either extensions or value[x], not both extension.exists() != value.exists()
obs-6 error Observation dataAbsentReason SHALL only be present if Observation.value[x] is not present dataAbsentReason.empty() or value.empty()
obs-7 error Observation If Observation.component.code is the same as Observation.code, then Observation.value SHALL NOT be present (the Observation.component.value[x] holds the value). value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty()
obs-8 error Observation bodyStructure SHALL only be present if Observation.bodySite is not present bodySite.exists() implies bodyStructure.empty()

NameFlagsCard.TypeDescription & Constraints    Filter: Filtersdoco
.. Observation 0..* Observation Measurements and simple assertions
... Slices for extension 1..* Extension Extension
Slice: Unordered, Open by value:url
.... extension:ciedDevice S 1..* Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) Cardiac implantable device(s)
URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension
... status S 1..1 code registered | preliminary | final | amended +
... code S 1..1 CodeableConcept Type of observation (code / type)
Required Pattern: At least the following
.... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
..... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
..... code 1..1 code Symbol in syntax defined by the system
Fixed Value: observation-cied-connectivity
... effective[x] S 1..1 dateTime Clinically relevant time/time-period for observation
... device S 0..1 Reference(Device - CIED Monitor Profile) A reference to the device that generates the measurements or the device settings for the device
... Slices for component S 1..* BackboneElement CIED Connectivity Components
Slice: Unordered, Open by value:code
.... component:All Slices Content/Rules for all slices
..... code S 1..1 CodeableConcept Type of component observation (code / type)
..... value[x] S 0..1 Quantity, CodeableConcept, string, boolean, integer, Range, Ratio, SampledData, time, dateTime, Period, Attachment, Reference(MolecularSequence) Actual component result
.... component:connectivityStatus S 1..1 BackboneElement The overall system connectivity status as calculated by the CIED manufacturer.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-status
..... value[x] 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Value Set (required)
.... component:connectivityModifier S 0..* BackboneElement Additional modifiers to connectivity status as calculated by the CIED manufacturer.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-modifier
..... value[x] 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Modifier Value Set (required)
.... component:statusReason S 0..1 BackboneElement Text describing reasons why the patient may be listed at a given status and/or modifier.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-reason
..... value[x] 0..1 string Actual component result
.... component:statusGuidance S 0..1 BackboneElement Text describing steps to get the patient to a connected status
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-guidance
..... value[x] 0..1 string Actual component result
.... component:lastCiedConnectivityDate S 0..1 BackboneElement The timestamp of the most recent communication between the CIED and monitor.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-cied-connectivity-time
..... value[x] 0..1 dateTime Actual component result
.... component:lastMonitorConnectivityDate S 0..1 BackboneElement The timestamp of the most recent communication from the monitor to the remote monitoring platform.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-monitor-connectivity-time
..... value[x] 0..1 dateTime Actual component result
.... component:nextCiedConnectivityDate S 0..1 BackboneElement The timestamp of the next expected communication between the CIED and monitor.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-cied-connectivity-date
..... value[x] 0..1 dateTime Actual component result
.... component:nextMonitorConnectivityDate S 0..1 BackboneElement The timestamp of the next expected communication from the monitor to the remote monitoring platform.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-monitor-connectivity-date
..... value[x] 0..1 dateTime Actual component result
.... component:lastRemoteInterrogationDate S 0..1 BackboneElement The date of the most recent remote device interrogation with a personal monitor.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-interrogation-date
..... value[x] 0..1 dateTime Actual component result
.... component:nextScheduledRemoteInterrogationDate S 0..1 BackboneElement The date of the next scheduled remote device interrogation.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-interrogation-date
..... value[x] 0..1 dateTime Actual component result
.... component:nextSecondaryScheduledRemoteInterrogationDate S 0..1 BackboneElement The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-secondary-interrogation-date
..... value[x] 0..1 dateTime Actual component result
.... component:connectionInterval S 0..1 BackboneElement The time interval in days between expected CIED and remote monitoring software communication.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connection-interval
..... value[x] 0..1 integer Actual component result
.... component:expectedTimePeriod S 0..1 BackboneElement The time period in days in which successful communication must take place for the connectivity status to remain connected.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: expected-time-period
..... value[x] 0..1 integer Actual component result

doco Documentation for this format

Terminology Bindings (Differential)

Path Status Usage ValueSet Version Source
Observation.component:connectivityStatus.​value[x] Base required CIED Connectivity Status Value Set 📦1.0.0 This IG
Observation.component:connectivityModifier.​value[x] Base required CIED Connectivity Status Modifier Value Set 📦1.0.0 This IG
NameFlagsCard.TypeDescription & Constraints    Filter: Filtersdoco
.. Observation C 0..* Observation Measurements and simple assertions
Constraints: obs-6, obs-7, obs-8
... id Σ 0..1 id Logical id of this artifact
... meta Σ 0..1 Meta Metadata about the resource
... implicitRules ?!Σ 0..1 uri A set of rules under which this content was created
... language 0..1 code Language of the resource content
Binding: AllLanguages (required): IETF language tag for a human language
Additional BindingsPurpose
CommonLanguages Starter
... text 0..1 Narrative Text summary of the resource, for human interpretation
This profile does not constrain the narrative in regard to content, language, or traceability to data elements
... contained 0..* Resource Contained, inline Resources
... Slices for extension 1..* Extension Extension
Slice: Unordered, Open by value:url
.... extension:ciedDevice S 1..* Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) Cardiac implantable device(s)
URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension
... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored
... identifier Σ 0..* Identifier Business Identifier for observation
... instantiates[x] Σ 0..1 canonical(ObservationDefinition), Reference(ObservationDefinition) Instantiates FHIR ObservationDefinition
... basedOn Σ 0..* Reference(CarePlan | DeviceRequest | ImmunizationRecommendation | MedicationRequest | NutritionOrder | ServiceRequest) Fulfills plan, proposal or order
... triggeredBy 0..* BackboneElement Triggering observation(s)
.... id 0..1 string Unique id for inter-element referencing
.... extension 0..* Extension Additional content defined by implementations
.... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
.... observation Σ 1..1 Reference(Observation) Triggering observation
.... type Σ 1..1 code reflex | repeat | re-run
Binding: TriggeredBytype (required): The type of TriggeredBy Observation.
.... reason 0..1 string Reason that the observation was triggered
... partOf Σ 0..* Reference(MedicationAdministration | MedicationDispense | MedicationStatement | Procedure | Immunization | ImagingStudy | GenomicStudy) Part of referenced event
... status ?!SΣ 1..1 code registered | preliminary | final | amended +
Binding: ObservationStatus (required): Codes providing the status of an observation.
... category 0..* CodeableConcept Classification of type of observation
Binding: ObservationCategoryCodes (preferred): Codes for high level observation categories.
... code SΣC 1..1 CodeableConcept Type of observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
.... id 0..1 string Unique id for inter-element referencing
.... extension 0..* Extension Additional content defined by implementations
.... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
..... version 0..1 string Version of the system - if relevant
..... code 1..1 code Symbol in syntax defined by the system
Fixed Value: observation-cied-connectivity
..... display 0..1 string Representation defined by the system
..... userSelected 0..1 boolean If this coding was chosen directly by the user
.... text 0..1 string Plain text representation of the concept
... subject SΣ 1..1 Reference(Patient - CIED Patient Profile) Who and/or what the observation is about
... focus Σ 0..* Reference(Resource) What the observation is about, when it is not about the subject of record
... encounter Σ 0..1 Reference(Encounter) Healthcare event during which this observation is made
... effective[x] SΣ 1..1 dateTime Clinically relevant time/time-period for observation
... issued Σ 0..1 instant Date/Time this version was made available
... performer Σ 0..* Reference(Practitioner | PractitionerRole | Organization | CareTeam | Patient | RelatedPerson) Who is responsible for the observation
... value[x] ΣC 0..1 Actual result
.... valueQuantity Quantity
.... valueCodeableConcept CodeableConcept
.... valueString string
.... valueBoolean boolean
.... valueInteger integer
.... valueRange Range
.... valueRatio Ratio
.... valueSampledData SampledData
.... valueTime time
.... valueDateTime dateTime
.... valuePeriod Period
.... valueAttachment Attachment
.... valueReference Reference(MolecularSequence)
... dataAbsentReason C 0..1 CodeableConcept Why the result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
... note 0..* Annotation Comments about the observation
... bodySite C 0..1 CodeableConcept Observed body part
Binding: SNOMEDCTBodyStructures (example): SNOMED CT Body site concepts
... bodyStructure C 0..1 Reference(BodyStructure) Observed body structure
... method 0..1 CodeableConcept How it was done
Binding: ObservationMethods (example): Methods for simple observations.
... specimen C 0..1 Reference(Specimen | Group) Specimen used for this observation
Constraints: obs-9
... device S 0..1 Reference(Device - CIED Monitor Profile) A reference to the device that generates the measurements or the device settings for the device
... referenceRange C 0..* BackboneElement Provides guide for interpretation
Constraints: obs-3
.... id 0..1 string Unique id for inter-element referencing
.... extension 0..* Extension Additional content defined by implementations
.... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
.... low C 0..1 SimpleQuantity(5.0.0) Low Range, if relevant
.... high C 0..1 SimpleQuantity(5.0.0) High Range, if relevant
.... normalValue 0..1 CodeableConcept Normal value, if relevant
Binding: ObservationReferenceRangeNormalValueCodes (extensible): Codes identifying the normal value of the observation.
.... type 0..1 CodeableConcept Reference range qualifier
Binding: ObservationReferenceRangeMeaningCodes (preferred): Code for the meaning of a reference range.
.... appliesTo 0..* CodeableConcept Reference range population
Binding: ObservationReferenceRangeAppliesToCodes (example): Codes identifying the population the reference range applies to.
.... age 0..1 Range Applicable age range, if relevant
.... text C 0..1 markdown Text based reference range in an observation
... hasMember Σ 0..* Reference(Observation | QuestionnaireResponse | MolecularSequence) Related resource that belongs to the Observation group
... derivedFrom Σ 0..* Reference(DocumentReference | ImagingStudy | ImagingSelection | QuestionnaireResponse | Observation | MolecularSequence | GenomicStudy) Related resource from which the observation is made
... Slices for component SΣC 1..* BackboneElement CIED Connectivity Components
Slice: Unordered, Open by value:code
.... component:All Slices Content/Rules for all slices
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
..... value[x] SΣ 0..1 Actual component result
...... valueQuantity Quantity
...... valueCodeableConcept CodeableConcept
...... valueString string
...... valueBoolean boolean
...... valueInteger integer
...... valueRange Range
...... valueRatio Ratio
...... valueSampledData SampledData
...... valueTime time
...... valueDateTime dateTime
...... valuePeriod Period
...... valueAttachment Attachment
...... valueReference Reference(MolecularSequence)
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:connectivityStatus SΣC 1..1 BackboneElement The overall system connectivity status as calculated by the CIED manufacturer.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-status
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Value Set (required)
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:connectivityModifier SΣC 0..* BackboneElement Additional modifiers to connectivity status as calculated by the CIED manufacturer.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-modifier
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Modifier Value Set (required)
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:statusReason SΣC 0..1 BackboneElement Text describing reasons why the patient may be listed at a given status and/or modifier.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-reason
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 string Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:statusGuidance SΣC 0..1 BackboneElement Text describing steps to get the patient to a connected status
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-guidance
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 string Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:lastCiedConnectivityDate SΣC 0..1 BackboneElement The timestamp of the most recent communication between the CIED and monitor.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-cied-connectivity-time
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:lastMonitorConnectivityDate SΣC 0..1 BackboneElement The timestamp of the most recent communication from the monitor to the remote monitoring platform.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-monitor-connectivity-time
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:nextCiedConnectivityDate SΣC 0..1 BackboneElement The timestamp of the next expected communication between the CIED and monitor.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-cied-connectivity-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:nextMonitorConnectivityDate SΣC 0..1 BackboneElement The timestamp of the next expected communication from the monitor to the remote monitoring platform.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-monitor-connectivity-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:lastRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the most recent remote device interrogation with a personal monitor.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:nextScheduledRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the next scheduled remote device interrogation.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:nextSecondaryScheduledRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-secondary-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:connectionInterval SΣC 0..1 BackboneElement The time interval in days between expected CIED and remote monitoring software communication.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connection-interval
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 integer Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:expectedTimePeriod SΣC 0..1 BackboneElement The time period in days in which successful communication must take place for the connectivity status to remain connected.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: expected-time-period
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 integer Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result

doco Documentation for this format

Terminology Bindings

Path Status Usage ValueSet Version Source
Observation.language Base required All Languages 📍5.0.0 FHIR Std.
Observation.triggeredBy.​type Base required triggered Bytype 📍5.0.0 FHIR Std.
Observation.status Base required Observation Status 📍5.0.0 FHIR Std.
Observation.category Base preferred Observation Category Codes 📍5.0.0 FHIR Std.
Observation.code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.bodySite Base example SNOMED CT Body Structures 📍5.0.0 FHIR Std.
Observation.method Base example Observation Methods 📍5.0.0 FHIR Std.
Observation.referenceRange.​normalValue Base extensible Observation Reference Range Normal Value Codes 📍5.0.0 FHIR Std.
Observation.referenceRange.​type Base preferred Observation Reference Range Meaning Codes 📍5.0.0 FHIR Std.
Observation.referenceRange.​appliesTo Base example Observation Reference Range Applies To Codes 📍5.0.0 FHIR Std.
Observation.component.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​value[x] Base required CIED Connectivity Status Value Set 📦1.0.0 This IG
Observation.component:connectivityStatus.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityModifier.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityModifier.​value[x] Base required CIED Connectivity Status Modifier Value Set 📦1.0.0 This IG
Observation.component:connectivityModifier.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:connectivityModifier.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:statusReason.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:statusReason.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:statusReason.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:statusGuidance.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:statusGuidance.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:statusGuidance.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:lastCiedConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastCiedConnectivityDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:lastCiedConnectivityDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:lastMonitorConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastMonitorConnectivityDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:lastMonitorConnectivityDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:nextCiedConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextCiedConnectivityDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:nextCiedConnectivityDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:nextMonitorConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextMonitorConnectivityDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:nextMonitorConnectivityDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:lastRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastRemoteInterrogationDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:lastRemoteInterrogationDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:nextScheduledRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextScheduledRemoteInterrogationDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:nextScheduledRemoteInterrogationDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:connectionInterval.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectionInterval.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:connectionInterval.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:expectedTimePeriod.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:expectedTimePeriod.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:expectedTimePeriod.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.

Constraints

Id Grade Path(s) Description Expression
dom-2 error Observation If the resource is contained in another resource, it SHALL NOT contain nested Resources contained.contained.empty()
dom-3 error Observation If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().ofType(canonical) | %resource.descendants().ofType(uri) | %resource.descendants().ofType(url))) or descendants().where(reference = '#').exists() or descendants().where(ofType(canonical) = '#').exists() or descendants().where(ofType(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
dom-4 error Observation If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
dom-5 error Observation If a resource is contained in another resource, it SHALL NOT have a security label contained.meta.security.empty()
dom-6 best practice Observation A resource should have narrative for robust management text.`div`.exists()
ele-1 error **ALL** elements All FHIR elements must have a @value or children hasValue() or (children().count() > id.count())
ext-1 error **ALL** extensions Must have either extensions or value[x], not both extension.exists() != value.exists()
obs-3 error Observation.referenceRange Must have at least a low or a high or text low.exists() or high.exists() or text.exists()
obs-6 error Observation dataAbsentReason SHALL only be present if Observation.value[x] is not present dataAbsentReason.empty() or value.empty()
obs-7 error Observation If Observation.component.code is the same as Observation.code, then Observation.value SHALL NOT be present (the Observation.component.value[x] holds the value). value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty()
obs-8 error Observation bodyStructure SHALL only be present if Observation.bodySite is not present bodySite.exists() implies bodyStructure.empty()
obs-9 error Observation.specimen If Observation.specimen is a reference to Group, the group can only have specimens (reference.resolve().exists() and reference.resolve() is Group) implies reference.resolve().member.entity.resolve().all($this is Specimen)

Summary

Mandatory: 6 elements
Must-Support: 22 elements

Structures

This structure refers to these other structures:

Extensions

This structure refers to these extensions:

Slices

This structure defines the following Slices:

  • The element 1 is sliced based on the value of Observation.component

Maturity: 2

Key Elements View

NameFlagsCard.TypeDescription & Constraints    Filter: Filtersdoco
.. Observation C 0..* Observation Measurements and simple assertions
Constraints: obs-6, obs-7, obs-8
... implicitRules ?!Σ 0..1 uri A set of rules under which this content was created
... contained 0..* Resource Contained, inline Resources
... Slices for extension 1..* Extension Extension
Slice: Unordered, Open by value:url
.... extension:ciedDevice S 1..* Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) Cardiac implantable device(s)
URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension
... status ?!SΣ 1..1 code registered | preliminary | final | amended +
Binding: ObservationStatus (required): Codes providing the status of an observation.
... code SΣC 1..1 CodeableConcept Type of observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
.... id 0..1 string Unique id for inter-element referencing
.... extension 0..* Extension Additional content defined by implementations
.... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
..... version 0..1 string Version of the system - if relevant
..... code 1..1 code Symbol in syntax defined by the system
Fixed Value: observation-cied-connectivity
..... display 0..1 string Representation defined by the system
..... userSelected 0..1 boolean If this coding was chosen directly by the user
.... text 0..1 string Plain text representation of the concept
... subject SΣ 1..1 Reference(Patient - CIED Patient Profile) Who and/or what the observation is about
... effective[x] SΣ 1..1 dateTime Clinically relevant time/time-period for observation
... value[x] ΣC 0..1 Actual result
.... valueQuantity Quantity
.... valueCodeableConcept CodeableConcept
.... valueString string
.... valueBoolean boolean
.... valueInteger integer
.... valueRange Range
.... valueRatio Ratio
.... valueSampledData SampledData
.... valueTime time
.... valueDateTime dateTime
.... valuePeriod Period
.... valueAttachment Attachment
.... valueReference Reference(MolecularSequence)
... device S 0..1 Reference(Device - CIED Monitor Profile) A reference to the device that generates the measurements or the device settings for the device
... Slices for component SΣC 1..* BackboneElement CIED Connectivity Components
Slice: Unordered, Open by value:code
.... component:All Slices Content/Rules for all slices
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
..... value[x] SΣ 0..1 Actual component result
...... valueQuantity Quantity
...... valueCodeableConcept CodeableConcept
...... valueString string
...... valueBoolean boolean
...... valueInteger integer
...... valueRange Range
...... valueRatio Ratio
...... valueSampledData SampledData
...... valueTime time
...... valueDateTime dateTime
...... valuePeriod Period
...... valueAttachment Attachment
...... valueReference Reference(MolecularSequence)
.... component:connectivityStatus SΣC 1..1 BackboneElement The overall system connectivity status as calculated by the CIED manufacturer.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-status
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Value Set (required)
.... component:connectivityModifier SΣC 0..* BackboneElement Additional modifiers to connectivity status as calculated by the CIED manufacturer.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-modifier
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Modifier Value Set (required)
.... component:statusReason SΣC 0..1 BackboneElement Text describing reasons why the patient may be listed at a given status and/or modifier.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-reason
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 string Actual component result
.... component:statusGuidance SΣC 0..1 BackboneElement Text describing steps to get the patient to a connected status
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-guidance
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 string Actual component result
.... component:lastCiedConnectivityDate SΣC 0..1 BackboneElement The timestamp of the most recent communication between the CIED and monitor.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-cied-connectivity-time
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:lastMonitorConnectivityDate SΣC 0..1 BackboneElement The timestamp of the most recent communication from the monitor to the remote monitoring platform.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-monitor-connectivity-time
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:nextCiedConnectivityDate SΣC 0..1 BackboneElement The timestamp of the next expected communication between the CIED and monitor.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-cied-connectivity-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:nextMonitorConnectivityDate SΣC 0..1 BackboneElement The timestamp of the next expected communication from the monitor to the remote monitoring platform.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-monitor-connectivity-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:lastRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the most recent remote device interrogation with a personal monitor.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:nextScheduledRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the next scheduled remote device interrogation.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:nextSecondaryScheduledRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-secondary-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
.... component:connectionInterval SΣC 0..1 BackboneElement The time interval in days between expected CIED and remote monitoring software communication.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connection-interval
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 integer Actual component result
.... component:expectedTimePeriod SΣC 0..1 BackboneElement The time period in days in which successful communication must take place for the connectivity status to remain connected.
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: expected-time-period
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 integer Actual component result

doco Documentation for this format

Terminology Bindings

Path Status Usage ValueSet Version Source
Observation.status Base required Observation Status 📍5.0.0 FHIR Std.
Observation.code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​value[x] Base required CIED Connectivity Status Value Set 📦1.0.0 This IG
Observation.component:connectivityModifier.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityModifier.​value[x] Base required CIED Connectivity Status Modifier Value Set 📦1.0.0 This IG
Observation.component:statusReason.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:statusGuidance.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastCiedConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastMonitorConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextCiedConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextMonitorConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextScheduledRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectionInterval.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:expectedTimePeriod.​code Base example LOINC Codes 📍5.0.0 FHIR Std.

Constraints

Id Grade Path(s) Description Expression
dom-2 error Observation If the resource is contained in another resource, it SHALL NOT contain nested Resources contained.contained.empty()
dom-3 error Observation If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().ofType(canonical) | %resource.descendants().ofType(uri) | %resource.descendants().ofType(url))) or descendants().where(reference = '#').exists() or descendants().where(ofType(canonical) = '#').exists() or descendants().where(ofType(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
dom-4 error Observation If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
dom-5 error Observation If a resource is contained in another resource, it SHALL NOT have a security label contained.meta.security.empty()
dom-6 best practice Observation A resource should have narrative for robust management text.`div`.exists()
ele-1 error **ALL** elements All FHIR elements must have a @value or children hasValue() or (children().count() > id.count())
ext-1 error **ALL** extensions Must have either extensions or value[x], not both extension.exists() != value.exists()
obs-6 error Observation dataAbsentReason SHALL only be present if Observation.value[x] is not present dataAbsentReason.empty() or value.empty()
obs-7 error Observation If Observation.component.code is the same as Observation.code, then Observation.value SHALL NOT be present (the Observation.component.value[x] holds the value). value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty()
obs-8 error Observation bodyStructure SHALL only be present if Observation.bodySite is not present bodySite.exists() implies bodyStructure.empty()

Differential View

NameFlagsCard.TypeDescription & Constraints    Filter: Filtersdoco
.. Observation 0..* Observation Measurements and simple assertions
... Slices for extension 1..* Extension Extension
Slice: Unordered, Open by value:url
.... extension:ciedDevice S 1..* Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) Cardiac implantable device(s)
URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension
... status S 1..1 code registered | preliminary | final | amended +
... code S 1..1 CodeableConcept Type of observation (code / type)
Required Pattern: At least the following
.... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
..... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
..... code 1..1 code Symbol in syntax defined by the system
Fixed Value: observation-cied-connectivity
... effective[x] S 1..1 dateTime Clinically relevant time/time-period for observation
... device S 0..1 Reference(Device - CIED Monitor Profile) A reference to the device that generates the measurements or the device settings for the device
... Slices for component S 1..* BackboneElement CIED Connectivity Components
Slice: Unordered, Open by value:code
.... component:All Slices Content/Rules for all slices
..... code S 1..1 CodeableConcept Type of component observation (code / type)
..... value[x] S 0..1 Quantity, CodeableConcept, string, boolean, integer, Range, Ratio, SampledData, time, dateTime, Period, Attachment, Reference(MolecularSequence) Actual component result
.... component:connectivityStatus S 1..1 BackboneElement The overall system connectivity status as calculated by the CIED manufacturer.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-status
..... value[x] 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Value Set (required)
.... component:connectivityModifier S 0..* BackboneElement Additional modifiers to connectivity status as calculated by the CIED manufacturer.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-modifier
..... value[x] 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Modifier Value Set (required)
.... component:statusReason S 0..1 BackboneElement Text describing reasons why the patient may be listed at a given status and/or modifier.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-reason
..... value[x] 0..1 string Actual component result
.... component:statusGuidance S 0..1 BackboneElement Text describing steps to get the patient to a connected status
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-guidance
..... value[x] 0..1 string Actual component result
.... component:lastCiedConnectivityDate S 0..1 BackboneElement The timestamp of the most recent communication between the CIED and monitor.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-cied-connectivity-time
..... value[x] 0..1 dateTime Actual component result
.... component:lastMonitorConnectivityDate S 0..1 BackboneElement The timestamp of the most recent communication from the monitor to the remote monitoring platform.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-monitor-connectivity-time
..... value[x] 0..1 dateTime Actual component result
.... component:nextCiedConnectivityDate S 0..1 BackboneElement The timestamp of the next expected communication between the CIED and monitor.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-cied-connectivity-date
..... value[x] 0..1 dateTime Actual component result
.... component:nextMonitorConnectivityDate S 0..1 BackboneElement The timestamp of the next expected communication from the monitor to the remote monitoring platform.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-monitor-connectivity-date
..... value[x] 0..1 dateTime Actual component result
.... component:lastRemoteInterrogationDate S 0..1 BackboneElement The date of the most recent remote device interrogation with a personal monitor.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-interrogation-date
..... value[x] 0..1 dateTime Actual component result
.... component:nextScheduledRemoteInterrogationDate S 0..1 BackboneElement The date of the next scheduled remote device interrogation.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-interrogation-date
..... value[x] 0..1 dateTime Actual component result
.... component:nextSecondaryScheduledRemoteInterrogationDate S 0..1 BackboneElement The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-secondary-interrogation-date
..... value[x] 0..1 dateTime Actual component result
.... component:connectionInterval S 0..1 BackboneElement The time interval in days between expected CIED and remote monitoring software communication.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connection-interval
..... value[x] 0..1 integer Actual component result
.... component:expectedTimePeriod S 0..1 BackboneElement The time period in days in which successful communication must take place for the connectivity status to remain connected.
..... code 1..1 CodeableConcept Type of component observation (code / type)
Required Pattern: At least the following
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: expected-time-period
..... value[x] 0..1 integer Actual component result

doco Documentation for this format

Terminology Bindings (Differential)

Path Status Usage ValueSet Version Source
Observation.component:connectivityStatus.​value[x] Base required CIED Connectivity Status Value Set 📦1.0.0 This IG
Observation.component:connectivityModifier.​value[x] Base required CIED Connectivity Status Modifier Value Set 📦1.0.0 This IG

Snapshot ViewView

NameFlagsCard.TypeDescription & Constraints    Filter: Filtersdoco
.. Observation C 0..* Observation Measurements and simple assertions
Constraints: obs-6, obs-7, obs-8
... id Σ 0..1 id Logical id of this artifact
... meta Σ 0..1 Meta Metadata about the resource
... implicitRules ?!Σ 0..1 uri A set of rules under which this content was created
... language 0..1 code Language of the resource content
Binding: AllLanguages (required): IETF language tag for a human language
Additional BindingsPurpose
CommonLanguages Starter
... text 0..1 Narrative Text summary of the resource, for human interpretation
This profile does not constrain the narrative in regard to content, language, or traceability to data elements
... contained 0..* Resource Contained, inline Resources
... Slices for extension 1..* Extension Extension
Slice: Unordered, Open by value:url
.... extension:ciedDevice S 1..* Reference(Device - Cardiovascular Implantable Electronic Device (CIED)) Cardiac implantable device(s)
URL: http://hl7.org/fhir/uv/cardx-cied/StructureDefinition/cied-device-extension
... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored
... identifier Σ 0..* Identifier Business Identifier for observation
... instantiates[x] Σ 0..1 canonical(ObservationDefinition), Reference(ObservationDefinition) Instantiates FHIR ObservationDefinition
... basedOn Σ 0..* Reference(CarePlan | DeviceRequest | ImmunizationRecommendation | MedicationRequest | NutritionOrder | ServiceRequest) Fulfills plan, proposal or order
... triggeredBy 0..* BackboneElement Triggering observation(s)
.... id 0..1 string Unique id for inter-element referencing
.... extension 0..* Extension Additional content defined by implementations
.... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
.... observation Σ 1..1 Reference(Observation) Triggering observation
.... type Σ 1..1 code reflex | repeat | re-run
Binding: TriggeredBytype (required): The type of TriggeredBy Observation.
.... reason 0..1 string Reason that the observation was triggered
... partOf Σ 0..* Reference(MedicationAdministration | MedicationDispense | MedicationStatement | Procedure | Immunization | ImagingStudy | GenomicStudy) Part of referenced event
... status ?!SΣ 1..1 code registered | preliminary | final | amended +
Binding: ObservationStatus (required): Codes providing the status of an observation.
... category 0..* CodeableConcept Classification of type of observation
Binding: ObservationCategoryCodes (preferred): Codes for high level observation categories.
... code SΣC 1..1 CodeableConcept Type of observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
.... id 0..1 string Unique id for inter-element referencing
.... extension 0..* Extension Additional content defined by implementations
.... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
..... version 0..1 string Version of the system - if relevant
..... code 1..1 code Symbol in syntax defined by the system
Fixed Value: observation-cied-connectivity
..... display 0..1 string Representation defined by the system
..... userSelected 0..1 boolean If this coding was chosen directly by the user
.... text 0..1 string Plain text representation of the concept
... subject SΣ 1..1 Reference(Patient - CIED Patient Profile) Who and/or what the observation is about
... focus Σ 0..* Reference(Resource) What the observation is about, when it is not about the subject of record
... encounter Σ 0..1 Reference(Encounter) Healthcare event during which this observation is made
... effective[x] SΣ 1..1 dateTime Clinically relevant time/time-period for observation
... issued Σ 0..1 instant Date/Time this version was made available
... performer Σ 0..* Reference(Practitioner | PractitionerRole | Organization | CareTeam | Patient | RelatedPerson) Who is responsible for the observation
... value[x] ΣC 0..1 Actual result
.... valueQuantity Quantity
.... valueCodeableConcept CodeableConcept
.... valueString string
.... valueBoolean boolean
.... valueInteger integer
.... valueRange Range
.... valueRatio Ratio
.... valueSampledData SampledData
.... valueTime time
.... valueDateTime dateTime
.... valuePeriod Period
.... valueAttachment Attachment
.... valueReference Reference(MolecularSequence)
... dataAbsentReason C 0..1 CodeableConcept Why the result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
... note 0..* Annotation Comments about the observation
... bodySite C 0..1 CodeableConcept Observed body part
Binding: SNOMEDCTBodyStructures (example): SNOMED CT Body site concepts
... bodyStructure C 0..1 Reference(BodyStructure) Observed body structure
... method 0..1 CodeableConcept How it was done
Binding: ObservationMethods (example): Methods for simple observations.
... specimen C 0..1 Reference(Specimen | Group) Specimen used for this observation
Constraints: obs-9
... device S 0..1 Reference(Device - CIED Monitor Profile) A reference to the device that generates the measurements or the device settings for the device
... referenceRange C 0..* BackboneElement Provides guide for interpretation
Constraints: obs-3
.... id 0..1 string Unique id for inter-element referencing
.... extension 0..* Extension Additional content defined by implementations
.... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
.... low C 0..1 SimpleQuantity(5.0.0) Low Range, if relevant
.... high C 0..1 SimpleQuantity(5.0.0) High Range, if relevant
.... normalValue 0..1 CodeableConcept Normal value, if relevant
Binding: ObservationReferenceRangeNormalValueCodes (extensible): Codes identifying the normal value of the observation.
.... type 0..1 CodeableConcept Reference range qualifier
Binding: ObservationReferenceRangeMeaningCodes (preferred): Code for the meaning of a reference range.
.... appliesTo 0..* CodeableConcept Reference range population
Binding: ObservationReferenceRangeAppliesToCodes (example): Codes identifying the population the reference range applies to.
.... age 0..1 Range Applicable age range, if relevant
.... text C 0..1 markdown Text based reference range in an observation
... hasMember Σ 0..* Reference(Observation | QuestionnaireResponse | MolecularSequence) Related resource that belongs to the Observation group
... derivedFrom Σ 0..* Reference(DocumentReference | ImagingStudy | ImagingSelection | QuestionnaireResponse | Observation | MolecularSequence | GenomicStudy) Related resource from which the observation is made
... Slices for component SΣC 1..* BackboneElement CIED Connectivity Components
Slice: Unordered, Open by value:code
.... component:All Slices Content/Rules for all slices
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
..... value[x] SΣ 0..1 Actual component result
...... valueQuantity Quantity
...... valueCodeableConcept CodeableConcept
...... valueString string
...... valueBoolean boolean
...... valueInteger integer
...... valueRange Range
...... valueRatio Ratio
...... valueSampledData SampledData
...... valueTime time
...... valueDateTime dateTime
...... valuePeriod Period
...... valueAttachment Attachment
...... valueReference Reference(MolecularSequence)
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:connectivityStatus SΣC 1..1 BackboneElement The overall system connectivity status as calculated by the CIED manufacturer.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-status
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Value Set (required)
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:connectivityModifier SΣC 0..* BackboneElement Additional modifiers to connectivity status as calculated by the CIED manufacturer.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connectivity-modifier
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 CodeableConcept Actual component result
Binding: CIED Connectivity Status Modifier Value Set (required)
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:statusReason SΣC 0..1 BackboneElement Text describing reasons why the patient may be listed at a given status and/or modifier.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-reason
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 string Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:statusGuidance SΣC 0..1 BackboneElement Text describing steps to get the patient to a connected status
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: status-guidance
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 string Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:lastCiedConnectivityDate SΣC 0..1 BackboneElement The timestamp of the most recent communication between the CIED and monitor.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-cied-connectivity-time
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:lastMonitorConnectivityDate SΣC 0..1 BackboneElement The timestamp of the most recent communication from the monitor to the remote monitoring platform.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-monitor-connectivity-time
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:nextCiedConnectivityDate SΣC 0..1 BackboneElement The timestamp of the next expected communication between the CIED and monitor.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-cied-connectivity-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:nextMonitorConnectivityDate SΣC 0..1 BackboneElement The timestamp of the next expected communication from the monitor to the remote monitoring platform.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-monitor-connectivity-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:lastRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the most recent remote device interrogation with a personal monitor.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: last-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:nextScheduledRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the next scheduled remote device interrogation.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:nextSecondaryScheduledRemoteInterrogationDate SΣC 0..1 BackboneElement The date of the next scheduled remote device interrogation by the secondary clinic for patients who are dual-managed.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: next-secondary-interrogation-date
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 dateTime Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:connectionInterval SΣC 0..1 BackboneElement The time interval in days between expected CIED and remote monitoring software communication.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: connection-interval
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 integer Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result
.... component:expectedTimePeriod SΣC 0..1 BackboneElement The time period in days in which successful communication must take place for the connectivity status to remain connected.
..... id 0..1 string Unique id for inter-element referencing
..... extension 0..* Extension Additional content defined by implementations
..... modifierExtension ?!Σ 0..* Extension Extensions that cannot be ignored even if unrecognized
..... code SΣC 1..1 CodeableConcept Type of component observation (code / type)
Binding: LOINCCodes (example): Codes identifying names of simple observations.
Required Pattern: At least the following
...... id 0..1 string Unique id for inter-element referencing
...... extension 0..* Extension Additional content defined by implementations
...... coding 1..* Coding Code defined by a terminology system
Fixed Value: (Complex)
....... id 0..1 string Unique id for inter-element referencing
....... extension 0..* Extension Additional content defined by implementations
....... system 1..1 uri Identity of the terminology system
Fixed Value: http://hl7.org/fhir/uv/cardx-cied/CodeSystem/CardXCIED
....... version 0..1 string Version of the system - if relevant
....... code 1..1 code Symbol in syntax defined by the system
Fixed Value: expected-time-period
....... display 0..1 string Representation defined by the system
....... userSelected 0..1 boolean If this coding was chosen directly by the user
...... text 0..1 string Plain text representation of the concept
..... value[x] SΣ 0..1 integer Actual component result
..... dataAbsentReason 0..1 CodeableConcept Why the component result is missing
Binding: DataAbsentReason (extensible): Codes specifying why the result (Observation.value[x]) is missing.
..... interpretation 0..* CodeableConcept High, low, normal, etc
Binding: ObservationInterpretationCodes (extensible): Codes identifying interpretations of observations.
..... referenceRange 0..* See referenceRange (Observation) Provides guide for interpretation of component result

doco Documentation for this format

Terminology Bindings

Path Status Usage ValueSet Version Source
Observation.language Base required All Languages 📍5.0.0 FHIR Std.
Observation.triggeredBy.​type Base required triggered Bytype 📍5.0.0 FHIR Std.
Observation.status Base required Observation Status 📍5.0.0 FHIR Std.
Observation.category Base preferred Observation Category Codes 📍5.0.0 FHIR Std.
Observation.code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.bodySite Base example SNOMED CT Body Structures 📍5.0.0 FHIR Std.
Observation.method Base example Observation Methods 📍5.0.0 FHIR Std.
Observation.referenceRange.​normalValue Base extensible Observation Reference Range Normal Value Codes 📍5.0.0 FHIR Std.
Observation.referenceRange.​type Base preferred Observation Reference Range Meaning Codes 📍5.0.0 FHIR Std.
Observation.referenceRange.​appliesTo Base example Observation Reference Range Applies To Codes 📍5.0.0 FHIR Std.
Observation.component.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​value[x] Base required CIED Connectivity Status Value Set 📦1.0.0 This IG
Observation.component:connectivityStatus.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:connectivityStatus.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityModifier.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectivityModifier.​value[x] Base required CIED Connectivity Status Modifier Value Set 📦1.0.0 This IG
Observation.component:connectivityModifier.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:connectivityModifier.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:statusReason.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:statusReason.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:statusReason.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:statusGuidance.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:statusGuidance.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:statusGuidance.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:lastCiedConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastCiedConnectivityDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:lastCiedConnectivityDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:lastMonitorConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastMonitorConnectivityDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:lastMonitorConnectivityDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:nextCiedConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextCiedConnectivityDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:nextCiedConnectivityDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:nextMonitorConnectivityDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextMonitorConnectivityDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:nextMonitorConnectivityDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:lastRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:lastRemoteInterrogationDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:lastRemoteInterrogationDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:nextScheduledRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextScheduledRemoteInterrogationDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:nextScheduledRemoteInterrogationDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:nextSecondaryScheduledRemoteInterrogationDate.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:connectionInterval.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:connectionInterval.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:connectionInterval.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.
Observation.component:expectedTimePeriod.​code Base example LOINC Codes 📍5.0.0 FHIR Std.
Observation.component:expectedTimePeriod.​dataAbsentReason Base extensible Data Absent Reason 📍5.0.0 FHIR Std.
Observation.component:expectedTimePeriod.​interpretation Base extensible Observation Interpretation Codes 📍5.0.0 FHIR Std.

Constraints

Id Grade Path(s) Description Expression
dom-2 error Observation If the resource is contained in another resource, it SHALL NOT contain nested Resources contained.contained.empty()
dom-3 error Observation If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().ofType(canonical) | %resource.descendants().ofType(uri) | %resource.descendants().ofType(url))) or descendants().where(reference = '#').exists() or descendants().where(ofType(canonical) = '#').exists() or descendants().where(ofType(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
dom-4 error Observation If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
dom-5 error Observation If a resource is contained in another resource, it SHALL NOT have a security label contained.meta.security.empty()
dom-6 best practice Observation A resource should have narrative for robust management text.`div`.exists()
ele-1 error **ALL** elements All FHIR elements must have a @value or children hasValue() or (children().count() > id.count())
ext-1 error **ALL** extensions Must have either extensions or value[x], not both extension.exists() != value.exists()
obs-3 error Observation.referenceRange Must have at least a low or a high or text low.exists() or high.exists() or text.exists()
obs-6 error Observation dataAbsentReason SHALL only be present if Observation.value[x] is not present dataAbsentReason.empty() or value.empty()
obs-7 error Observation If Observation.component.code is the same as Observation.code, then Observation.value SHALL NOT be present (the Observation.component.value[x] holds the value). value.empty() or component.code.where(coding.intersect(%resource.code.coding).exists()).empty()
obs-8 error Observation bodyStructure SHALL only be present if Observation.bodySite is not present bodySite.exists() implies bodyStructure.empty()
obs-9 error Observation.specimen If Observation.specimen is a reference to Group, the group can only have specimens (reference.resolve().exists() and reference.resolve() is Group) implies reference.resolve().member.entity.resolve().all($this is Specimen)

Summary

Mandatory: 6 elements
Must-Support: 22 elements

Structures

This structure refers to these other structures:

Extensions

This structure refers to these extensions:

Slices

This structure defines the following Slices:

  • The element 1 is sliced based on the value of Observation.component

Maturity: 2

 

Other representations of profile: CSV, Excel, Schematron