Personal Health Device Implementation Guide
2.0.1 - STU 2 International flag

Personal Health Device Implementation Guide, published by HL7 International / Devices. This guide is not an authorized publication; it is the continuous build for version 2.0.1 built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/phd/ and changes regularly. See the Directory of published versions

Resource Profile: PhdDeviceMetric

Official URL: http://hl7.org/fhir/uv/phd/StructureDefinition/PhdDeviceMetric Version: 2.0.1
Standards status: Trial-use Maturity Level: 2 Computable Name: PhdDeviceMetric
Other Identifiers: OID:2.16.840.1.113883.4.642.40.73.42.11

Profile for the DeviceMetric for a PHD

ACOM and GHS provide the list of observation types that a device can generate. Accuracy can be specified as well. In GHS the valid range and schedule information can be present. This information can be mapped to a PhdDeviceMetric FHIR Resource as follows:

Usages:

You can also check for usages in the FHIR IG Statistics

Formal Views of Profile Content

Description Differentials, Snapshots, and other representations.

NameFlagsCard.TypeDescription & Constraints    Filter: doco
DeviceMetric 0..* DeviceMetric Measurement, calculation or setting capability of a medical device
DeviceMetric.implicitRules This element is a modifier element. This element is included in summaries. 0..1 uri A set of rules under which this content was created
DeviceMetric.Slices for extension 0..* Extension Extension
Slice: Unordered, Open by value:url

Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.Slices for extension.extension:Accuracy 0..1 Quantity Numeric Accuracy
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Accuracy
DeviceMetric.Slices for extension.extension:Confidence95 0..1 Range Confidence95
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Confidence95
DeviceMetric.Slices for extension.extension:SimpleAlerting 0..1 (Complex) SimpleAlerting
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/SimpleAlerting
DeviceMetric.modifierExtension This element is a modifier element. 0..* Extension Extensions that cannot be ignored
DeviceMetric.type This element is included in summaries. 1..1 CodeableConcept Identity of metric, for example Heart Rate or PEEP Setting
Binding: DeviceMetricAndComponentTypes (preferred): Describes the metric type.
DeviceMetric.type.Slices for coding This element is included in summaries. 1..* Coding Code defined by a terminology system
Slice: Unordered, Open by value:system

Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.type.Slices for coding.coding:All Slices Content/Rules for all slices
DeviceMetric.type.Slices for coding.coding:All Slices.system This element is included in summaries. 0..1 uri Identity of the terminology system
DeviceMetric.type.Slices for coding.coding:All Slices.code This element is included in summaries. 0..1 code Symbol in syntax defined by the system
DeviceMetric.type.Slices for coding.coding:MDCType This element is included in summaries. 1..1 Coding The 11073-10101 MDC code for the metric
DeviceMetric.type.Slices for coding.coding:MDCType.system This element is included in summaries. 1..1 uri Identity of the terminology system
Fixed Value: urn:iso:std:iso:11073:10101
DeviceMetric.type.Slices for coding.coding:MDCType.code This element is included in summaries. 1..1 code Symbol in syntax defined by the system
DeviceMetric.type.text This element is included in summaries. 0..1 string It is recommended to display at least the reference identifier for the MDC code.
DeviceMetric.unit This element is included in summaries. 0..1 CodeableConcept Unit of Measure for the Metric
Binding: DeviceMetricAndComponentTypes (preferred): Describes the unit of the metric.
DeviceMetric.source This element is included in summaries. 1..1 Reference(PhdDevice) The device that is producing the metric
DeviceMetric.operationalStatus This element is included in summaries. 0..1 code on | off | standby | entered-in-error
Binding: DeviceMetricOperationalStatus (required): Describes the operational status of the DeviceMetric.
DeviceMetric.category This element is included in summaries. 1..1 code measurement | setting | calculation | unspecified
Binding: DeviceMetricCategory (required): Describes the category of the metric.
DeviceMetric.measurementPeriod This element is included in summaries. 0..1 Timing Describes the measurement repetition time

doco Documentation for this format

Terminology Bindings

Path Status Usage ValueSet Version Source
DeviceMetric.type Base preferred Device Metric and Component Types 📍4.0.1 FHIR Std.
DeviceMetric.unit Base preferred Device Metric and Component Types 📍4.0.1 FHIR Std.
DeviceMetric.operationalStatus Base required DeviceMetricOperationalStatus 📍4.0.1 FHIR Std.
DeviceMetric.category Base required DeviceMetricCategory 📍4.0.1 FHIR Std.

Constraints

Id Grade Path(s) Description Expression
dom-2 error DeviceMetric If the resource is contained in another resource, it SHALL NOT contain nested Resources contained.contained.empty()
dom-3 error DeviceMetric If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().as(canonical) | %resource.descendants().as(uri) | %resource.descendants().as(url))) or descendants().where(reference = '#').exists() or descendants().where(as(canonical) = '#').exists() or descendants().where(as(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
dom-4 error DeviceMetric If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
dom-5 error DeviceMetric If a resource is contained in another resource, it SHALL NOT have a security label contained.meta.security.empty()
dom-6 best practice DeviceMetric A resource should have narrative for robust management text.`div`.exists()
ele-1 error **ALL** elements All FHIR elements must have a @value or children hasValue() or (children().count() > id.count())
ext-1 error **ALL** extensions Must have either extensions or value[x], not both extension.exists() != value.exists()

This structure is derived from DeviceMetric

NameFlagsCard.TypeDescription & Constraints    Filter: doco
DeviceMetric 0..* DeviceMetric Measurement, calculation or setting capability of a medical device
DeviceMetric.Slices for extension 0..* Extension Extension
Slice: Unordered, Open by value:url
Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.Slices for extension.extension:Accuracy 0..1 Quantity Numeric Accuracy
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Accuracy
DeviceMetric.Slices for extension.extension:Confidence95 0..1 Range Confidence95
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Confidence95
DeviceMetric.Slices for extension.extension:SimpleAlerting 0..1 (Complex) SimpleAlerting
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/SimpleAlerting
DeviceMetric.type
DeviceMetric.type.Slices for coding 1..* Coding Code defined by a terminology system
Slice: Unordered, Open by value:system
Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.type.Slices for coding.coding:All Slices Content/Rules for all slices
DeviceMetric.type.Slices for coding.coding:All Slices.system 0..1 uri Identity of the terminology system
DeviceMetric.type.Slices for coding.coding:All Slices.code 0..1 code Symbol in syntax defined by the system
DeviceMetric.type.Slices for coding.coding:MDCType 1..1 Coding The 11073-10101 MDC code for the metric
DeviceMetric.type.Slices for coding.coding:MDCType.system 1..1 uri Identity of the terminology system
Fixed Value: urn:iso:std:iso:11073:10101
DeviceMetric.type.Slices for coding.coding:MDCType.code 1..1 code Symbol in syntax defined by the system
DeviceMetric.type.text 0..1 string It is recommended to display at least the reference identifier for the MDC code.
DeviceMetric.unit 0..1 CodeableConcept Unit of Measure for the Metric
DeviceMetric.parent 0..0 No reference to a parent device in PHD. PHDs are not part of a device hierarchy.
DeviceMetric.operationalStatus 0..1 code on | off | standby | entered-in-error
DeviceMetric.category 1..1 code measurement | setting | calculation | unspecified
DeviceMetric.measurementPeriod 0..1 Timing Describes the measurement repetition time

doco Documentation for this format
NameFlagsCard.TypeDescription & Constraints    Filter: doco
DeviceMetric 0..* DeviceMetric Measurement, calculation or setting capability of a medical device
DeviceMetric.id This element is included in summaries. 0..1 id Logical id of this artifact
DeviceMetric.meta This element is included in summaries. 0..1 Meta Metadata about the resource
DeviceMetric.implicitRules This element is a modifier element. This element is included in summaries. 0..1 uri A set of rules under which this content was created
DeviceMetric.language 0..1 code Language of the resource content
Binding: CommonLanguages (preferred): A human language.
Additional BindingsPurpose
AllLanguages Max Binding
DeviceMetric.text 0..1 Narrative Text summary of the resource, for human interpretation
This profile does not constrain the narrative in regard to content, language, or traceability to data elements
DeviceMetric.contained 0..* Resource Contained, inline Resources
DeviceMetric.Slices for extension 0..* Extension Extension
Slice: Unordered, Open by value:url

Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.Slices for extension.extension:Accuracy 0..1 Quantity Numeric Accuracy
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Accuracy
DeviceMetric.Slices for extension.extension:Confidence95 0..1 Range Confidence95
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Confidence95
DeviceMetric.Slices for extension.extension:SimpleAlerting 0..1 (Complex) SimpleAlerting
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/SimpleAlerting
DeviceMetric.modifierExtension This element is a modifier element. 0..* Extension Extensions that cannot be ignored
DeviceMetric.identifier This element is included in summaries. 0..* Identifier Instance identifier
DeviceMetric.type This element is included in summaries. 1..1 CodeableConcept Identity of metric, for example Heart Rate or PEEP Setting
Binding: DeviceMetricAndComponentTypes (preferred): Describes the metric type.
DeviceMetric.type.id 0..1 string Unique id for inter-element referencing
DeviceMetric.type.extension 0..* Extension Additional content defined by implementations
Slice: Unordered, Open by value:url
DeviceMetric.type.Slices for coding This element is included in summaries. 1..* Coding Code defined by a terminology system
Slice: Unordered, Open by value:system

Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.type.Slices for coding.coding:All Slices Content/Rules for all slices
DeviceMetric.type.Slices for coding.coding:All Slices.id 0..1 string Unique id for inter-element referencing
DeviceMetric.type.Slices for coding.coding:All Slices.extension 0..* Extension Additional content defined by implementations
Slice: Unordered, Open by value:url
DeviceMetric.type.Slices for coding.coding:All Slices.system This element is included in summaries. 0..1 uri Identity of the terminology system
DeviceMetric.type.Slices for coding.coding:All Slices.version This element is included in summaries. 0..1 string Version of the system - if relevant
DeviceMetric.type.Slices for coding.coding:All Slices.code This element is included in summaries. 0..1 code Symbol in syntax defined by the system
DeviceMetric.type.Slices for coding.coding:All Slices.display This element is included in summaries. 0..1 string Representation defined by the system
DeviceMetric.type.Slices for coding.coding:All Slices.userSelected This element is included in summaries. 0..1 boolean If this coding was chosen directly by the user
DeviceMetric.type.Slices for coding.coding:MDCType This element is included in summaries. 1..1 Coding The 11073-10101 MDC code for the metric
DeviceMetric.type.Slices for coding.coding:MDCType.id 0..1 string Unique id for inter-element referencing
DeviceMetric.type.Slices for coding.coding:MDCType.extension 0..* Extension Additional content defined by implementations
Slice: Unordered, Open by value:url
DeviceMetric.type.Slices for coding.coding:MDCType.system This element is included in summaries. 1..1 uri Identity of the terminology system
Fixed Value: urn:iso:std:iso:11073:10101
DeviceMetric.type.Slices for coding.coding:MDCType.version This element is included in summaries. 0..1 string Version of the system - if relevant
DeviceMetric.type.Slices for coding.coding:MDCType.code This element is included in summaries. 1..1 code Symbol in syntax defined by the system
DeviceMetric.type.Slices for coding.coding:MDCType.display This element is included in summaries. 0..1 string Representation defined by the system
DeviceMetric.type.Slices for coding.coding:MDCType.userSelected This element is included in summaries. 0..1 boolean If this coding was chosen directly by the user
DeviceMetric.type.text This element is included in summaries. 0..1 string It is recommended to display at least the reference identifier for the MDC code.
DeviceMetric.unit This element is included in summaries. 0..1 CodeableConcept Unit of Measure for the Metric
Binding: DeviceMetricAndComponentTypes (preferred): Describes the unit of the metric.
DeviceMetric.source This element is included in summaries. 1..1 Reference(PhdDevice) The device that is producing the metric
DeviceMetric.operationalStatus This element is included in summaries. 0..1 code on | off | standby | entered-in-error
Binding: DeviceMetricOperationalStatus (required): Describes the operational status of the DeviceMetric.
DeviceMetric.color This element is included in summaries. 0..1 code black | red | green | yellow | blue | magenta | cyan | white
Binding: DeviceMetricColor (required): Describes the typical color of representation.
DeviceMetric.category This element is included in summaries. 1..1 code measurement | setting | calculation | unspecified
Binding: DeviceMetricCategory (required): Describes the category of the metric.
DeviceMetric.measurementPeriod This element is included in summaries. 0..1 Timing Describes the measurement repetition time
DeviceMetric.calibration This element is included in summaries. 0..* BackboneElement Describes the calibrations that have been performed or that are required to be performed
DeviceMetric.calibration.id 0..1 string Unique id for inter-element referencing
DeviceMetric.calibration.extension 0..* Extension Additional content defined by implementations
DeviceMetric.calibration.modifierExtension This element is a modifier element. This element is included in summaries. 0..* Extension Extensions that cannot be ignored even if unrecognized
DeviceMetric.calibration.type This element is included in summaries. 0..1 code unspecified | offset | gain | two-point
Binding: DeviceMetricCalibrationType (required): Describes the type of a metric calibration.
DeviceMetric.calibration.state This element is included in summaries. 0..1 code not-calibrated | calibration-required | calibrated | unspecified
Binding: DeviceMetricCalibrationState (required): Describes the state of a metric calibration.
DeviceMetric.calibration.time This element is included in summaries. 0..1 instant Describes the time last calibration has been performed

doco Documentation for this format

Terminology Bindings

Path Status Usage ValueSet Version Source
DeviceMetric.language Base preferred Common Languages 📍4.0.1 FHIR Std.
DeviceMetric.type Base preferred Device Metric and Component Types 📍4.0.1 FHIR Std.
DeviceMetric.unit Base preferred Device Metric and Component Types 📍4.0.1 FHIR Std.
DeviceMetric.operationalStatus Base required DeviceMetricOperationalStatus 📍4.0.1 FHIR Std.
DeviceMetric.color Base required DeviceMetricColor 📍4.0.1 FHIR Std.
DeviceMetric.category Base required DeviceMetricCategory 📍4.0.1 FHIR Std.
DeviceMetric.calibration.​type Base required DeviceMetricCalibrationType 📍4.0.1 FHIR Std.
DeviceMetric.calibration.​state Base required DeviceMetricCalibrationState 📍4.0.1 FHIR Std.

Constraints

Id Grade Path(s) Description Expression
dom-2 error DeviceMetric If the resource is contained in another resource, it SHALL NOT contain nested Resources contained.contained.empty()
dom-3 error DeviceMetric If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().as(canonical) | %resource.descendants().as(uri) | %resource.descendants().as(url))) or descendants().where(reference = '#').exists() or descendants().where(as(canonical) = '#').exists() or descendants().where(as(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
dom-4 error DeviceMetric If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
dom-5 error DeviceMetric If a resource is contained in another resource, it SHALL NOT have a security label contained.meta.security.empty()
dom-6 best practice DeviceMetric A resource should have narrative for robust management text.`div`.exists()
ele-1 error **ALL** elements All FHIR elements must have a @value or children hasValue() or (children().count() > id.count())
ext-1 error **ALL** extensions Must have either extensions or value[x], not both extension.exists() != value.exists()

This structure is derived from DeviceMetric

Summary

Mandatory: 5 elements
Fixed: 1 element
Prohibited: 1 element

Structures

This structure refers to these other structures:

Extensions

This structure refers to these extensions:

Slices

This structure defines the following Slices:

  • The element 1 is sliced based on the value of DeviceMetric.type.coding

Maturity: 2

Key Elements View

NameFlagsCard.TypeDescription & Constraints    Filter: doco
DeviceMetric 0..* DeviceMetric Measurement, calculation or setting capability of a medical device
DeviceMetric.implicitRules This element is a modifier element. This element is included in summaries. 0..1 uri A set of rules under which this content was created
DeviceMetric.Slices for extension 0..* Extension Extension
Slice: Unordered, Open by value:url

Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.Slices for extension.extension:Accuracy 0..1 Quantity Numeric Accuracy
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Accuracy
DeviceMetric.Slices for extension.extension:Confidence95 0..1 Range Confidence95
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Confidence95
DeviceMetric.Slices for extension.extension:SimpleAlerting 0..1 (Complex) SimpleAlerting
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/SimpleAlerting
DeviceMetric.modifierExtension This element is a modifier element. 0..* Extension Extensions that cannot be ignored
DeviceMetric.type This element is included in summaries. 1..1 CodeableConcept Identity of metric, for example Heart Rate or PEEP Setting
Binding: DeviceMetricAndComponentTypes (preferred): Describes the metric type.
DeviceMetric.type.Slices for coding This element is included in summaries. 1..* Coding Code defined by a terminology system
Slice: Unordered, Open by value:system

Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.type.Slices for coding.coding:All Slices Content/Rules for all slices
DeviceMetric.type.Slices for coding.coding:All Slices.system This element is included in summaries. 0..1 uri Identity of the terminology system
DeviceMetric.type.Slices for coding.coding:All Slices.code This element is included in summaries. 0..1 code Symbol in syntax defined by the system
DeviceMetric.type.Slices for coding.coding:MDCType This element is included in summaries. 1..1 Coding The 11073-10101 MDC code for the metric
DeviceMetric.type.Slices for coding.coding:MDCType.system This element is included in summaries. 1..1 uri Identity of the terminology system
Fixed Value: urn:iso:std:iso:11073:10101
DeviceMetric.type.Slices for coding.coding:MDCType.code This element is included in summaries. 1..1 code Symbol in syntax defined by the system
DeviceMetric.type.text This element is included in summaries. 0..1 string It is recommended to display at least the reference identifier for the MDC code.
DeviceMetric.unit This element is included in summaries. 0..1 CodeableConcept Unit of Measure for the Metric
Binding: DeviceMetricAndComponentTypes (preferred): Describes the unit of the metric.
DeviceMetric.source This element is included in summaries. 1..1 Reference(PhdDevice) The device that is producing the metric
DeviceMetric.operationalStatus This element is included in summaries. 0..1 code on | off | standby | entered-in-error
Binding: DeviceMetricOperationalStatus (required): Describes the operational status of the DeviceMetric.
DeviceMetric.category This element is included in summaries. 1..1 code measurement | setting | calculation | unspecified
Binding: DeviceMetricCategory (required): Describes the category of the metric.
DeviceMetric.measurementPeriod This element is included in summaries. 0..1 Timing Describes the measurement repetition time

doco Documentation for this format

Terminology Bindings

Path Status Usage ValueSet Version Source
DeviceMetric.type Base preferred Device Metric and Component Types 📍4.0.1 FHIR Std.
DeviceMetric.unit Base preferred Device Metric and Component Types 📍4.0.1 FHIR Std.
DeviceMetric.operationalStatus Base required DeviceMetricOperationalStatus 📍4.0.1 FHIR Std.
DeviceMetric.category Base required DeviceMetricCategory 📍4.0.1 FHIR Std.

Constraints

Id Grade Path(s) Description Expression
dom-2 error DeviceMetric If the resource is contained in another resource, it SHALL NOT contain nested Resources contained.contained.empty()
dom-3 error DeviceMetric If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().as(canonical) | %resource.descendants().as(uri) | %resource.descendants().as(url))) or descendants().where(reference = '#').exists() or descendants().where(as(canonical) = '#').exists() or descendants().where(as(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
dom-4 error DeviceMetric If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
dom-5 error DeviceMetric If a resource is contained in another resource, it SHALL NOT have a security label contained.meta.security.empty()
dom-6 best practice DeviceMetric A resource should have narrative for robust management text.`div`.exists()
ele-1 error **ALL** elements All FHIR elements must have a @value or children hasValue() or (children().count() > id.count())
ext-1 error **ALL** extensions Must have either extensions or value[x], not both extension.exists() != value.exists()

Differential View

This structure is derived from DeviceMetric

NameFlagsCard.TypeDescription & Constraints    Filter: doco
DeviceMetric 0..* DeviceMetric Measurement, calculation or setting capability of a medical device
DeviceMetric.Slices for extension 0..* Extension Extension
Slice: Unordered, Open by value:url
Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.Slices for extension.extension:Accuracy 0..1 Quantity Numeric Accuracy
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Accuracy
DeviceMetric.Slices for extension.extension:Confidence95 0..1 Range Confidence95
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Confidence95
DeviceMetric.Slices for extension.extension:SimpleAlerting 0..1 (Complex) SimpleAlerting
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/SimpleAlerting
DeviceMetric.type
DeviceMetric.type.Slices for coding 1..* Coding Code defined by a terminology system
Slice: Unordered, Open by value:system
Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.type.Slices for coding.coding:All Slices Content/Rules for all slices
DeviceMetric.type.Slices for coding.coding:All Slices.system 0..1 uri Identity of the terminology system
DeviceMetric.type.Slices for coding.coding:All Slices.code 0..1 code Symbol in syntax defined by the system
DeviceMetric.type.Slices for coding.coding:MDCType 1..1 Coding The 11073-10101 MDC code for the metric
DeviceMetric.type.Slices for coding.coding:MDCType.system 1..1 uri Identity of the terminology system
Fixed Value: urn:iso:std:iso:11073:10101
DeviceMetric.type.Slices for coding.coding:MDCType.code 1..1 code Symbol in syntax defined by the system
DeviceMetric.type.text 0..1 string It is recommended to display at least the reference identifier for the MDC code.
DeviceMetric.unit 0..1 CodeableConcept Unit of Measure for the Metric
DeviceMetric.source 1..1 Reference(PhdDevice) The device that is producing the metric
DeviceMetric.parent 0..0 No reference to a parent device in PHD. PHDs are not part of a device hierarchy.
DeviceMetric.operationalStatus 0..1 code on | off | standby | entered-in-error
DeviceMetric.category 1..1 code measurement | setting | calculation | unspecified
DeviceMetric.measurementPeriod 0..1 Timing Describes the measurement repetition time

doco Documentation for this format

Snapshot ViewView

NameFlagsCard.TypeDescription & Constraints    Filter: doco
DeviceMetric 0..* DeviceMetric Measurement, calculation or setting capability of a medical device
DeviceMetric.id This element is included in summaries. 0..1 id Logical id of this artifact
DeviceMetric.meta This element is included in summaries. 0..1 Meta Metadata about the resource
DeviceMetric.implicitRules This element is a modifier element. This element is included in summaries. 0..1 uri A set of rules under which this content was created
DeviceMetric.language 0..1 code Language of the resource content
Binding: CommonLanguages (preferred): A human language.
Additional BindingsPurpose
AllLanguages Max Binding
DeviceMetric.text 0..1 Narrative Text summary of the resource, for human interpretation
This profile does not constrain the narrative in regard to content, language, or traceability to data elements
DeviceMetric.contained 0..* Resource Contained, inline Resources
DeviceMetric.Slices for extension 0..* Extension Extension
Slice: Unordered, Open by value:url

Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.Slices for extension.extension:Accuracy 0..1 Quantity Numeric Accuracy
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Accuracy
DeviceMetric.Slices for extension.extension:Confidence95 0..1 Range Confidence95
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/Confidence95
DeviceMetric.Slices for extension.extension:SimpleAlerting 0..1 (Complex) SimpleAlerting
URL: http://hl7.org/fhir/uv/phd/StructureDefinition/SimpleAlerting
DeviceMetric.modifierExtension This element is a modifier element. 0..* Extension Extensions that cannot be ignored
DeviceMetric.identifier This element is included in summaries. 0..* Identifier Instance identifier
DeviceMetric.type This element is included in summaries. 1..1 CodeableConcept Identity of metric, for example Heart Rate or PEEP Setting
Binding: DeviceMetricAndComponentTypes (preferred): Describes the metric type.
DeviceMetric.type.id 0..1 string Unique id for inter-element referencing
DeviceMetric.type.extension 0..* Extension Additional content defined by implementations
Slice: Unordered, Open by value:url
DeviceMetric.type.Slices for coding This element is included in summaries. 1..* Coding Code defined by a terminology system
Slice: Unordered, Open by value:system

Slicing Note: The rules on this element (including its children) apply to all the slices, but are not repeated in them
DeviceMetric.type.Slices for coding.coding:All Slices Content/Rules for all slices
DeviceMetric.type.Slices for coding.coding:All Slices.id 0..1 string Unique id for inter-element referencing
DeviceMetric.type.Slices for coding.coding:All Slices.extension 0..* Extension Additional content defined by implementations
Slice: Unordered, Open by value:url
DeviceMetric.type.Slices for coding.coding:All Slices.system This element is included in summaries. 0..1 uri Identity of the terminology system
DeviceMetric.type.Slices for coding.coding:All Slices.version This element is included in summaries. 0..1 string Version of the system - if relevant
DeviceMetric.type.Slices for coding.coding:All Slices.code This element is included in summaries. 0..1 code Symbol in syntax defined by the system
DeviceMetric.type.Slices for coding.coding:All Slices.display This element is included in summaries. 0..1 string Representation defined by the system
DeviceMetric.type.Slices for coding.coding:All Slices.userSelected This element is included in summaries. 0..1 boolean If this coding was chosen directly by the user
DeviceMetric.type.Slices for coding.coding:MDCType This element is included in summaries. 1..1 Coding The 11073-10101 MDC code for the metric
DeviceMetric.type.Slices for coding.coding:MDCType.id 0..1 string Unique id for inter-element referencing
DeviceMetric.type.Slices for coding.coding:MDCType.extension 0..* Extension Additional content defined by implementations
Slice: Unordered, Open by value:url
DeviceMetric.type.Slices for coding.coding:MDCType.system This element is included in summaries. 1..1 uri Identity of the terminology system
Fixed Value: urn:iso:std:iso:11073:10101
DeviceMetric.type.Slices for coding.coding:MDCType.version This element is included in summaries. 0..1 string Version of the system - if relevant
DeviceMetric.type.Slices for coding.coding:MDCType.code This element is included in summaries. 1..1 code Symbol in syntax defined by the system
DeviceMetric.type.Slices for coding.coding:MDCType.display This element is included in summaries. 0..1 string Representation defined by the system
DeviceMetric.type.Slices for coding.coding:MDCType.userSelected This element is included in summaries. 0..1 boolean If this coding was chosen directly by the user
DeviceMetric.type.text This element is included in summaries. 0..1 string It is recommended to display at least the reference identifier for the MDC code.
DeviceMetric.unit This element is included in summaries. 0..1 CodeableConcept Unit of Measure for the Metric
Binding: DeviceMetricAndComponentTypes (preferred): Describes the unit of the metric.
DeviceMetric.source This element is included in summaries. 1..1 Reference(PhdDevice) The device that is producing the metric
DeviceMetric.operationalStatus This element is included in summaries. 0..1 code on | off | standby | entered-in-error
Binding: DeviceMetricOperationalStatus (required): Describes the operational status of the DeviceMetric.
DeviceMetric.color This element is included in summaries. 0..1 code black | red | green | yellow | blue | magenta | cyan | white
Binding: DeviceMetricColor (required): Describes the typical color of representation.
DeviceMetric.category This element is included in summaries. 1..1 code measurement | setting | calculation | unspecified
Binding: DeviceMetricCategory (required): Describes the category of the metric.
DeviceMetric.measurementPeriod This element is included in summaries. 0..1 Timing Describes the measurement repetition time
DeviceMetric.calibration This element is included in summaries. 0..* BackboneElement Describes the calibrations that have been performed or that are required to be performed
DeviceMetric.calibration.id 0..1 string Unique id for inter-element referencing
DeviceMetric.calibration.extension 0..* Extension Additional content defined by implementations
DeviceMetric.calibration.modifierExtension This element is a modifier element. This element is included in summaries. 0..* Extension Extensions that cannot be ignored even if unrecognized
DeviceMetric.calibration.type This element is included in summaries. 0..1 code unspecified | offset | gain | two-point
Binding: DeviceMetricCalibrationType (required): Describes the type of a metric calibration.
DeviceMetric.calibration.state This element is included in summaries. 0..1 code not-calibrated | calibration-required | calibrated | unspecified
Binding: DeviceMetricCalibrationState (required): Describes the state of a metric calibration.
DeviceMetric.calibration.time This element is included in summaries. 0..1 instant Describes the time last calibration has been performed

doco Documentation for this format

Terminology Bindings

Path Status Usage ValueSet Version Source
DeviceMetric.language Base preferred Common Languages 📍4.0.1 FHIR Std.
DeviceMetric.type Base preferred Device Metric and Component Types 📍4.0.1 FHIR Std.
DeviceMetric.unit Base preferred Device Metric and Component Types 📍4.0.1 FHIR Std.
DeviceMetric.operationalStatus Base required DeviceMetricOperationalStatus 📍4.0.1 FHIR Std.
DeviceMetric.color Base required DeviceMetricColor 📍4.0.1 FHIR Std.
DeviceMetric.category Base required DeviceMetricCategory 📍4.0.1 FHIR Std.
DeviceMetric.calibration.​type Base required DeviceMetricCalibrationType 📍4.0.1 FHIR Std.
DeviceMetric.calibration.​state Base required DeviceMetricCalibrationState 📍4.0.1 FHIR Std.

Constraints

Id Grade Path(s) Description Expression
dom-2 error DeviceMetric If the resource is contained in another resource, it SHALL NOT contain nested Resources contained.contained.empty()
dom-3 error DeviceMetric If the resource is contained in another resource, it SHALL be referred to from elsewhere in the resource or SHALL refer to the containing resource contained.where((('#'+id in (%resource.descendants().reference | %resource.descendants().as(canonical) | %resource.descendants().as(uri) | %resource.descendants().as(url))) or descendants().where(reference = '#').exists() or descendants().where(as(canonical) = '#').exists() or descendants().where(as(canonical) = '#').exists()).not()).trace('unmatched', id).empty()
dom-4 error DeviceMetric If a resource is contained in another resource, it SHALL NOT have a meta.versionId or a meta.lastUpdated contained.meta.versionId.empty() and contained.meta.lastUpdated.empty()
dom-5 error DeviceMetric If a resource is contained in another resource, it SHALL NOT have a security label contained.meta.security.empty()
dom-6 best practice DeviceMetric A resource should have narrative for robust management text.`div`.exists()
ele-1 error **ALL** elements All FHIR elements must have a @value or children hasValue() or (children().count() > id.count())
ext-1 error **ALL** extensions Must have either extensions or value[x], not both extension.exists() != value.exists()

This structure is derived from DeviceMetric

Summary

Mandatory: 5 elements
Fixed: 1 element
Prohibited: 1 element

Structures

This structure refers to these other structures:

Extensions

This structure refers to these extensions:

Slices

This structure defines the following Slices:

  • The element 1 is sliced based on the value of DeviceMetric.type.coding

Maturity: 2

 

Other representations of profile: CSV, Excel, Schematron