Personal Health Device Implementation Guide, published by HL7 International / Devices. This guide is not an authorized publication; it is the continuous build for version 3.0.0-ballot built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/phd/ and changes regularly. See the Directory of published versions

Example DeviceMetric: Phd Device Metric for Pulse Rate with UCUM

Page standards status: Informative

Why unit.text is used

This example uses unit.text for the human-readable label "beats per minute" while unit.coding carries the machine-readable unit codes. The UCUM coding uses /min and intentionally omits Coding.display: when a UCUM display is supplied, it must be a canonical UCUM display (such as /min), not a custom phrase like "beats per minute". Keeping the familiar label in unit.text avoids a UCUM display-name validation error while still making the unit understandable to readers.

The second coding, MDC code 2720 (MDC_DIM_BEAT_PER_MIN), provides the corresponding IEEE 11073 unit code. The MDC coding has its own canonical display and does not change the UCUM display requirement.