CH LAB-Order (R4)
3.0.0-ci-build - ci-build
CH LAB-Order (R4), published by HL7 Switzerland. This guide is not an authorized publication; it is the continuous build for version 3.0.0-ci-build built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/hl7ch/ch-lab-order/ and changes regularly. See the Directory of published versions
Generated Narrative: SpecimenDefinition specimen-serum-plasma
Language: en-US
Profile: Laboratory IVD Specimen Definition
typeCollected: Blood venous
collection: Venipuncture for blood test
typeTested
type: Serum
preference: Preferred
Containers
Material Type Cap Description Capacity MinimumVolume[x] plastic Evacuated blood collection tube with clot activator red cap red cap plastic tube 5 mL 10 milliliter (Details: UCUM codemL = 'mL') 5 milliliter (Details: UCUM codemL = 'mL') RetentionTimes
Value Unit System Code 24 hour http://unitsofmeasure.org h rejectionCriterion: blood specimen hemolized, insufficient quantity of specimen
handling
temperatureQualifier: room temperature or refrigerated
MaxDurations
Value Unit System Code 7 day http://unitsofmeasure.org d handling
temperatureQualifier: frozen
MaxDurations
Value Unit System Code 28 day http://unitsofmeasure.org d
typeTested
type: Plasma
preference: Alternate
Containers
Type Cap Description Capacity MinimumVolume[x] Evacuated blood collection tube with heparin lithium and gel separator green cap green cap plastic tube 5 mL 10 milliliter (Details: UCUM codemL = 'mL') 5 milliliter (Details: UCUM codemL = 'mL') RetentionTimes
Value Unit System Code 24 hour http://unitsofmeasure.org h rejectionCriterion: blood specimen hemolized, insufficient quantity of specimen
handling
temperatureQualifier: room temperature or refrigerated
MaxDurations
Value Unit System Code 7 day http://unitsofmeasure.org d handling
temperatureQualifier: frozen
MaxDurations
Value Unit System Code 28 day http://unitsofmeasure.org d