Point-of-Care Device Implementation Guide, published by HL7 International / Devices. This guide is not an authorized publication; it is the continuous build for version 0.4.0 built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/uv-pocd/ and changes regularly. See the Directory of published versions
| Page standards status: Informative |
<DeviceMetric xmlns="http://hl7.org/fhir">
<id value="HeartRateNumericDeviceMetric"/>
<meta>
<profile
value="http://hl7.org/fhir/uv/pocd/StructureDefinition/NumericDeviceMetric"/>
</meta>
<text>
<status value="extensions"/>
<div xmlns="http://www.w3.org/1999/xhtml"><p class="res-header-id"><b>Generated Narrative: DeviceMetric HeartRateNumericDeviceMetric</b></p><a name="HeartRateNumericDeviceMetric"> </a><a name="hcHeartRateNumericDeviceMetric"> </a><div style="display: inline-block; background-color: #d9e0e7; padding: 6px; margin: 4px; border: 1px solid #8da1b4; border-radius: 5px; line-height: 60%"><p style="margin-bottom: 0px"/><p style="margin-bottom: 0px">Profile: <a href="StructureDefinition-NumericDeviceMetric.html">Numeric DeviceMetric profile</a></p></div><blockquote><p><b>Relation extension</b></p><ul><li>kind: DCE</li><li>entries: <a href="DeviceMetric-ECGSampleArrayDeviceMetric.html">DeviceMetric: extension = 0.001,25 mm/s; type = MDC_ECG_ELEC_POTL; unit = MDC_DIM_MILLI_VOLT; operationalStatus = on; color = green; category = measurement</a></li></ul></blockquote><blockquote><p><b>Technical range extension</b></p><ul><li>lowerLimit: 0</li><li>upperLimit: 200</li><li>stepWidth: 1</li><li>relativeAccuracy: 0.01</li><li>absoluteAccuracy: 0.01</li></ul></blockquote><p><b>type</b>: <span title="Codes:{urn:iso:std:iso:11073:10101 147842}">HR</span></p><p><b>unit</b>: <span title="Codes:{urn:iso:std:iso:11073:10101 264864}">bpm</span></p><p><b>source</b>: <a href="Device-PhysioMonitorMdsDevice.html">Device: extension = ,->Practitioner Alice B. Careful ,8760,156,Dmo; identifier = urn:oid:1.2.840.10004.1.1.1.0.0.1.0.0.1.2680#01-23-45-67-89-AB-CD-EF; status = active; manufacturer = MyCompany; serialNumber = 123456; modelNumber = PM1; type = MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS</a></p><p><b>parent</b>: <a href="Device-HeartRateChannelDevice.html">Device: status = active; type = MDC_DEV_CARD_RATE_CHAN</a></p><p><b>operationalStatus</b>: On</p><p><b>color</b>: Color Green</p><p><b>category</b>: Measurement</p></div>
</text>
<extension url="http://hl7.org/fhir/uv/pocd/StructureDefinition/relation">
<extension url="kind">
<valueCode value="DCE"/>
</extension>
<extension url="entries">
<valueReference>🔗
<reference value="DeviceMetric/ECGSampleArrayDeviceMetric"/>
</valueReference>
</extension>
</extension>
<extension
url="http://hl7.org/fhir/uv/pocd/StructureDefinition/technical-range">
<extension url="lowerLimit">
<valueDecimal value="0"/>
</extension>
<extension url="upperLimit">
<valueDecimal value="200"/>
</extension>
<extension url="stepWidth">
<valueDecimal value="1"/>
</extension>
<extension url="relativeAccuracy">
<valueDecimal value="0.01"/>
</extension>
<extension url="absoluteAccuracy">
<valueDecimal value="0.01"/>
</extension>
</extension>
<type>
<coding>
<system value="urn:iso:std:iso:11073:10101"/>
<code value="147842"/>
<display value="MDC_ECG_HEART_RATE"/>
</coding>
<text value="HR"/>
</type>
<unit>
<coding>
<system value="urn:iso:std:iso:11073:10101"/>
<code value="264864"/>
<display value="MDC_DIM_BEAT_PER_MIN"/>
</coding>
<text value="bpm"/>
</unit>
<source>🔗
<reference value="Device/PhysioMonitorMdsDevice"/>
</source>
<parent>🔗
<reference value="Device/HeartRateChannelDevice"/>
</parent>
<operationalStatus value="on"/>
<color value="green"/>
<category value="measurement"/>
</DeviceMetric>