CH LAB-Order (R4)
2.0.0 - ci-build
CH LAB-Order (R4), published by HL7 Switzerland. This guide is not an authorized publication; it is the continuous build for version 2.0.0 built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/hl7ch/ch-lab-order/ and changes regularly. See the Directory of published versions
Generated Narrative: SpecimenDefinition
ResourceSpecimenDefinition "specimen-blood-EDTA" (Language"en-US")
Profile: Laboratory IVD Specimen Definition
typeCollected: Blood venous (specimenType#BLDV)
collection: Venipuncture for blood test (SNOMED CT#28520004)
typeTested
type: Whole blood (specimenType#BLD)
preference: preferred
Containers
Material Type Cap Description Capacity MinimumVolume[x] plastic (SNOMED CT#61088005 "Plastic") Blood tube (physical object) (SNOMED CT#706049005) red cap (ContainerCap#red) S-Monovette® EDTA 4.9 milliliter (Details: UCUM codemL = 'mL') 4 milliliter (Details: UCUM codemL = 'mL') RetentionTimes
Value Unit System Code 2 hour Unified Code for Units of Measure (UCUM) h rejectionCriterion: blood specimen hemolized (RejectionCriterion#hemolized), insufficient quantity of specimen (RejectionCriterion#insufficient)
handling
temperatureQualifier: room temperature or refrigerated ()
MaxDurations
Value Unit System Code 2 hour Unified Code for Units of Measure (UCUM) h