CH LAB-Order (R4), published by HL7 Switzerland. This guide is not an authorized publication; it is the continuous build for version 2.0.0 built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/hl7ch/ch-lab-order/ and changes regularly. See the Directory of published versions
Generated Narrative: SpecimenDefinition
ResourceSpecimenDefinition "single-test-serum-capillary-fasting"
typeCollected: Capillary blood specimen (SNOMED CT#122554006)
patientPreparation: 12 hour fasting (), At rest (SNOMED CT#263678003)
timeAspect: preferrably morning time
typeTested
type: Capillary blood specimen (SNOMED CT#122554006)
preference: preferred
Containers
Material Type Cap Description Capacity MinimumVolume[x] plastic (SNOMED CT#61088005) Evacuated blood collection tube with clot activator (SNOMED CT#767384004) red cap (ContainerCap#red) red cap plastic tube 2 mL 2 milliliter (Details: UCUM codemL = 'mL') 0.5 milliliter (Details: UCUM codemL = 'mL') handling
temperatureQualifier: Ambient temperature ()
temperatureRange: 15-25 °C
MaxDurations
Value Unit System Code 60 minute Unified Code for Units of Measure (UCUM) min handling
temperatureQualifier: Refrigerated temperature ()
temperatureRange: 2-8 °C
MaxDurations
Value Unit System Code 8 hour Unified Code for Units of Measure (UCUM) h
typeTested
type: Plasma specimen (SNOMED CT#119361006)
preference: alternate
Containers
Material Type Cap MinimumVolume[x] glass (SNOMED CT#32039001) Evacuated blood collection tube with heparin lithium and gel separator (SNOMED CT#767390000) green cap (ContainerCap#green) 2 ml (Details: UCUM codemL = 'mL') rejectionCriterion: insufficient specimen volume (RejectionCriterion#insufficient), hemolized specimen (RejectionCriterion#hemolized)
handling
temperatureQualifier: Ambient temperature ()
temperatureRange: 15-25 °C
MaxDurations
Value Unit System Code 60 minute Unified Code for Units of Measure (UCUM) min handling
temperatureQualifier: Refrigerated temperature ()
temperatureRange: 2-8 °C
MaxDurations
Value Unit System Code 8 hour Unified Code for Units of Measure (UCUM) h