Personal Health Device Implementation Guide
2.0.0 - STU 2 International flag

Personal Health Device Implementation Guide, published by HL7 International / Devices. This guide is not an authorized publication; it is the continuous build for version 2.0.0 built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/phd/ and changes regularly. See the Directory of published versions

ValueSet: MDC Clock Types

Official URL: http://hl7.org/fhir/uv/phd/ValueSet/MDCClockTypes Version: 2.0.0
Standards status: Trial-use Active as of 2021-09-25 Maturity Level: 2 Computable Name: MDCClockTypes

ValueSet for the ISO/IEEE 11073-10101 Nomenclature Codes for Clock Types

References

Logical Definition (CLD)

  • Include these codes as defined in urn:iso:std:iso:11073:10101 version ⏿2024-12-05
    CodeDisplayDefinition
    67975MDC_ATTR_TIME_ABSSpecifies the absolute date and time for system or device clock and derived objects (resolution of up to 1/100 of a second, encoded as eight pairs of 4-bit BCD nibbles denoting the { century, year, month, day, hour, minute, second, sec-fractions }).
    68226MDC_ATTR_TIME_STAMP_BOBase-offset time stamp value for observations and derived objects.
    67983MDC_ATTR_TIME_RELSpecifies the relative time value for system or device clock and derived objects (default 125 μs resolution counter having an arbitrary start time and value).
    68072MDC_ATTR_TIME_REL_HI_RESSpecifies the high-res relative time value for system or device clock and derived objects (default 1 μs resolution counter having an arbitrary start time and value).

 

Expansion

Expansion performed internally based on codesystem MDC Nomenclature v2024-12-05 (CodeSystem)

This value set contains 4 concepts

SystemCodeDisplay (en)DefinitionJSONXML
urn:iso:std:iso:11073:10101  67975MDC_ATTR_TIME_ABS

Specifies the absolute date and time for system or device clock and derived objects (resolution of up to 1/100 of a second, encoded as eight pairs of 4-bit BCD nibbles denoting the { century, year, month, day, hour, minute, second, sec-fractions }).

urn:iso:std:iso:11073:10101  68226MDC_ATTR_TIME_STAMP_BO

Base-offset time stamp value for observations and derived objects.

urn:iso:std:iso:11073:10101  67983MDC_ATTR_TIME_REL

Specifies the relative time value for system or device clock and derived objects (default 125 μs resolution counter having an arbitrary start time and value).

urn:iso:std:iso:11073:10101  68072MDC_ATTR_TIME_REL_HI_RES

Specifies the high-res relative time value for system or device clock and derived objects (default 1 μs resolution counter having an arbitrary start time and value).


Description of the above table(s).