Personal Health Device Implementation Guide, published by HL7 International / Devices. This guide is not an authorized publication; it is the continuous build for version 2.0.0 built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/phd/ and changes regularly. See the Directory of published versions
| Official URL: http://hl7.org/fhir/uv/phd/ValueSet/ASN1MeasurementBits | Version: 2.0.0 | ||||
| Standards status: Trial-use Active as of 2018-09-17 | Maturity Level: 2 | Computable Name: ASN1MeasurementBits | |||
ValueSet for the ANS1ToHL7 codes that are derived from enumeration measurements.
References
Profile: Shareable ValueSet
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 version 📦2.0.0 where source = measurement
Expansion performed internally based on codesystem ASN1 bits to HL7 enumerated values v2.0.0 (CodeSystem)
This value set contains 104 concepts
| System | Code | Display (en) | Definition | JSON | XML |
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 67846.0 | lim-alert-off | Both the high and the low limit alerts are disabled | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 67846.1 | lim-low-off | The low limit alert is disabled | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 67846.2 | lim-high-off | The high limit alert is disabled | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.0 | sensor-disconnected | The device is not connected with the sensor | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.1 | sensor-malfunction | The sensor has malfunctioned | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.2 | sensor-displaced | The sensor is not correctly placed on the user | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.3 | sensor-unsupported | The sensor connected to the device is unsupported | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.4 | sensor-off | The sensor is not connected to the user | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.5 | sensor-interference | The sensor is experiencing interference | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.6 | signal-searching | The sensor is searching for a signal | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.7 | signal-pulse-questionable | A questionable pulse has been detected | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.8 | signal-non-pulsatile | A non pulsatile signal has been detected | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.9 | signal-erratic | A signal has been discovered, but it is erratic | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.10 | signal-low-perfusion | The signal is experiencing low perfusion | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.11 | signal-poor | The signal has been discovered, but it is poor | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.12 | signal-inadequate | The signal has been discovered, but it is inadequate | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.13 | signal-processing-irregularity | The signal has been discovered, but is experiencing processing irregularity | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.14 | device-equipment-malfunction | A device malfunction has occurred | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150604.15 | device-extended-update | An extended display update is in progress on device | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150605.0 | pulse-qual-nominal | No abnormalities have been detected by the sensor | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150605.1 | pulse-qual-marginal | The pulse quality is marginal | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150605.2 | pulse-qual-minimal | The pulse quality is minimal | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 150605.3 | pulse-qual-unacceptable | The pulse quality is unacceptable | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410584.0 | leadwire-loss | Loss of unspecified lead wire or electrode connection | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410584.1 | leadsignal-loss | Loss of unspecified lead signal | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410584.2 | leadwire-loss-first-lead | Loss of 1st lead wire or electrode connection | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410584.3 | leadsignal-loss-first-lead | Loss of 1st lead signal | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410584.4 | leadwire-loss-second-lead | Loss of 2nd lead wire or electrode connection | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410584.5 | leadsignal-loss-second-lead | Loss of 2nd lead signal | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410584.6 | leadwire-loss-third-lead | Loss of 3rd lead wire or electrode connection | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410584.7 | leadsignal-loss-third-lead | Loss of 3rd lead signal | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.0 | device-battery-low | The device battery is low | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.1 | sensor-malfunction | The sensor has malfunctioned | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.2 | sensor-sample-size-insufficient | The sensor's sample size is insufficient | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.3 | sensor-strip-insertion | The test strip was inserted incorrectly | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.4 | sensor-strip-type-incorrect | The test strip type is incompatible with the sensor | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.5 | sensor-result-too-high | The value reported by the sensor is too high | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.6 | sensor-result-too-low | The value reported by the sensor is too low | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.7 | sensor-temp-too-high | The temperature is too high | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.8 | sensor-temp-too-low | The temperature is too low | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.9 | sensor-read-interrupt | The read process was interrupted | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.10 | device-gen-fault | General Device Fault | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417752.11 | sensor-temp-out-of-range | The temperature is out of range | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.0 | inr-device-battery-low | The INR device battery is low | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.1 | inr-sensor-malfunction | The INR device sensor has malfunctioned | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.2 | inr-sensor-sample-size-insufficient | The sample size was insufficient | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.3 | inr-sensor-strip-insertion | The test script was inserted incorrectly | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.4 | inr-sensor-strip-type-incorrect | The test strip used is not compatible with the INR | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.5 | inr-sensor-result-too-high | The test result is too high | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.6 | inr-sensor-result-too-low | The test result is too low | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.7 | inr-sensor-temp-too-high | The temperature is too high | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.8 | inr-sensor-temp-too-low | The temperature is too low | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.9 | inr-sensor-read-interrupt | The read process was interrupted | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.10 | inr-device-gen-fault | General device fault | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8417909.11 | inr-sensor-calibration-due | The INR sensor calibration is due | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.0 | device-status-undetermined | Undetermined device status event | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.1 | device-status-reset | Device reset | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.5 | device-status-error | Device error | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.6 | device-status-error-mechanical | Mechanical error | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.7 | device-status-error-electronic | Electronic error | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.8 | device-status-error-software | Software error | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.9 | device-status-error-battery | Battery error | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.15 | device-status-service | Undetermined device status event, service needed | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.16 | device-status-service-time-sync-required | Time sync service required | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.17 | device-status-service-calibration-required | Calibration required | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.18 | device-status-service-replenishment-required | Replenishment required | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.25 | device-status-battery-low | Battery power is too low | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.26 | device-status-battery-depleted | Battery is depleted | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.27 | device-status-battery-replaced | Battery has been replaced | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8408608.28 | device-status-battery-interrupted | Battery was interrupted | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.0 | sensor-session-stopped | Session stopped | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.2 | sensor-type-incorrect | Sensor type incorrect for device | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.3 | sensor-malfunction | The sensor malfunctioned | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.4 | device-specific-alert | Device specific alert | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.7 | sensor-calibration-not-allowed | Calibrating the sensor is not allowed | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.8 | sensor-calibration-recommended | Calibrating the sensor is recommended | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.9 | sensor-calibration-required | Calibrating the sensor is required | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.10 | sensor-temp-too-high | Temperature is too high for a valid result | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.11 | sensor-temp-too-low | Temperature is too low for a valid result | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.12 | sensor-result-below-patient-low | Sensor result below the Patient Low level | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.13 | sensor-result-above-patient-high | Sensor result above the Patient High level | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.14 | sensor-low-hypo | Sensor result below level | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.15 | sensor-high-hyper | Sensor result above level | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.16 | sensor-rate-decrease-exceeded | The sensor rate of decrease has been exceeded | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.17 | sensor-rate-increase-exceeded | The sensor rate of increase has been exceeded | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.18 | sensor-result-too-low | Sensor result is lower than the device can process | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.19 | sensor-result-too-high | Sensor result is higher than the device can process | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418060.20 | sensor-com-out-of-range | The sensor communication is out of range | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.0 | Battery-status-Undetermined | The status of the battery is unknown | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.1 | Battery-absent | The battery is absent | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.2 | Battery-active | The battery is active in the system | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.3 | Battery-charging | The battery is charging | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.4 | Battery-fullyCharged | The battery is fully charged | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.5 | Battery-disposable | The battery is disposable | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.6 | Battery-rechargeable | The battery is rechargeable | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.7 | Battery-overTemperature | The battery is too hot | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.8 | Battery-faulty | The battery is faulty | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8418512.9 | Battery-incompatible | The battery is incompatible | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410608.0 | body-movement | Too much body movement | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410608.1 | cuff-too-loose | The cuff was too loose | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410608.2 | irregular-pulse | An irregular pulse was detected | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410608.3 | pulse-over-range-limit | The pulse was too high for sensor to work | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410608.4 | pulse-under-range-limit | The pulse was too low for sensor to work | ||
http://terminology.hl7.org/CodeSystem/ASN1ToHL7 | 8410608.5 | improper-body-position | The cuff was incorrectly placed with respect to the heart |