Personal Health Device Implementation Guide, published by HL7 International / Devices. This guide is not an authorized publication; it is the continuous build for version 3.0.0-draft built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/phd/ and changes regularly. See the Directory of published versions

ValueSet: ANS1ToHL7 codes for measurement attributes

Official URL: http://hl7.org/fhir/uv/phd/ValueSet/ASN1MeasurementBits Version: 3.0.0-draft
Standards status: Trial-use Maturity Level: 1 Computable Name: ASN1MeasurementBits
Other Identifiers: OID:2.16.840.1.113883.4.642.40.73.48.4

ValueSet for the ANS1ToHL7 codes that are derived from enumeration measurements.

References

Logical Definition (CLD)

 

Expansion

Expansion performed internally based on codesystem ASN1 bits to HL7 enumerated values v3.0.0-draft (CodeSystem)

This value set contains 104 concepts

SystemCodeDisplay (en)DefinitionJSONXML
http://terminology.hl7.org/CodeSystem/ASN1ToHL7  67846.0lim-alert-off

Both the high and the low limit alerts are disabled

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  67846.1lim-low-off

The low limit alert is disabled

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  67846.2lim-high-off

The high limit alert is disabled

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.0sensor-disconnected

The device is not connected with the sensor

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.1sensor-malfunction

The sensor has malfunctioned

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.2sensor-displaced

The sensor is not correctly placed on the user

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.3sensor-unsupported

The sensor connected to the device is unsupported

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.4sensor-off

The sensor is not connected to the user

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.5sensor-interference

The sensor is experiencing interference

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.6signal-searching

The sensor is searching for a signal

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.7signal-pulse-questionable

A questionable pulse has been detected

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.8signal-non-pulsatile

A non pulsatile signal has been detected

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.9signal-erratic

A signal has been discovered, but it is erratic

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.10signal-low-perfusion

The signal is experiencing low perfusion

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.11signal-poor

The signal has been discovered, but it is poor

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.12signal-inadequate

The signal has been discovered, but it is inadequate

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.13signal-processing-irregularity

The signal has been discovered, but is experiencing processing irregularity

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.14device-equipment-malfunction

A device malfunction has occurred

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150604.15device-extended-update

An extended display update is in progress on device

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150605.0pulse-qual-nominal

No abnormalities have been detected by the sensor

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150605.1pulse-qual-marginal

The pulse quality is marginal

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150605.2pulse-qual-minimal

The pulse quality is minimal

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  150605.3pulse-qual-unacceptable

The pulse quality is unacceptable

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410584.0leadwire-loss

Loss of unspecified lead wire or electrode connection

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410584.1leadsignal-loss

Loss of unspecified lead signal

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410584.2leadwire-loss-first-lead

Loss of 1st lead wire or electrode connection

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410584.3leadsignal-loss-first-lead

Loss of 1st lead signal

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410584.4leadwire-loss-second-lead

Loss of 2nd lead wire or electrode connection

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410584.5leadsignal-loss-second-lead

Loss of 2nd lead signal

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410584.6leadwire-loss-third-lead

Loss of 3rd lead wire or electrode connection

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410584.7leadsignal-loss-third-lead

Loss of 3rd lead signal

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.0device-battery-low

The device battery is low

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.1sensor-malfunction

The sensor has malfunctioned

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.2sensor-sample-size-insufficient

The sensor's sample size is insufficient

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.3sensor-strip-insertion

The test strip was inserted incorrectly

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.4sensor-strip-type-incorrect

The test strip type is incompatible with the sensor

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.5sensor-result-too-high

The value reported by the sensor is too high

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.6sensor-result-too-low

The value reported by the sensor is too low

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.7sensor-temp-too-high

The temperature is too high

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.8sensor-temp-too-low

The temperature is too low

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.9sensor-read-interrupt

The read process was interrupted

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.10device-gen-fault

General Device Fault

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417752.11sensor-temp-out-of-range

The temperature is out of range

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.0inr-device-battery-low

The INR device battery is low

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.1inr-sensor-malfunction

The INR device sensor has malfunctioned

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.2inr-sensor-sample-size-insufficient

The sample size was insufficient

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.3inr-sensor-strip-insertion

The test script was inserted incorrectly

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.4inr-sensor-strip-type-incorrect

The test strip used is not compatible with the INR

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.5inr-sensor-result-too-high

The test result is too high

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.6inr-sensor-result-too-low

The test result is too low

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.7inr-sensor-temp-too-high

The temperature is too high

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.8inr-sensor-temp-too-low

The temperature is too low

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.9inr-sensor-read-interrupt

The read process was interrupted

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.10inr-device-gen-fault

General device fault

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8417909.11inr-sensor-calibration-due

The INR sensor calibration is due

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.0device-status-undetermined

Undetermined device status event

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.1device-status-reset

Device reset

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.5device-status-error

Device error

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.6device-status-error-mechanical

Mechanical error

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.7device-status-error-electronic

Electronic error

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.8device-status-error-software

Software error

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.9device-status-error-battery

Battery error

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.15device-status-service

Undetermined device status event, service needed

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.16device-status-service-time-sync-required

Time sync service required

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.17device-status-service-calibration-required

Calibration required

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.18device-status-service-replenishment-required

Replenishment required

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.25device-status-battery-low

Battery power is too low

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.26device-status-battery-depleted

Battery is depleted

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.27device-status-battery-replaced

Battery has been replaced

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8408608.28device-status-battery-interrupted

Battery was interrupted

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.0sensor-session-stopped

Session stopped

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.2sensor-type-incorrect

Sensor type incorrect for device

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.3sensor-malfunction

The sensor malfunctioned

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.4device-specific-alert

Device specific alert

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.7sensor-calibration-not-allowed

Calibrating the sensor is not allowed

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.8sensor-calibration-recommended

Calibrating the sensor is recommended

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.9sensor-calibration-required

Calibrating the sensor is required

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.10sensor-temp-too-high

Temperature is too high for a valid result

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.11sensor-temp-too-low

Temperature is too low for a valid result

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.12sensor-result-below-patient-low

Sensor result below the Patient Low level

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.13sensor-result-above-patient-high

Sensor result above the Patient High level

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.14sensor-low-hypo

Sensor result below level

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.15sensor-high-hyper

Sensor result above level

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.16sensor-rate-decrease-exceeded

The sensor rate of decrease has been exceeded

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.17sensor-rate-increase-exceeded

The sensor rate of increase has been exceeded

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.18sensor-result-too-low

Sensor result is lower than the device can process

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.19sensor-result-too-high

Sensor result is higher than the device can process

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418060.20sensor-com-out-of-range

The sensor communication is out of range

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.0Battery-status-Undetermined

The status of the battery is unknown

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.1Battery-absent

The battery is absent

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.2Battery-active

The battery is active in the system

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.3Battery-charging

The battery is charging

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.4Battery-fullyCharged

The battery is fully charged

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.5Battery-disposable

The battery is disposable

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.6Battery-rechargeable

The battery is rechargeable

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.7Battery-overTemperature

The battery is too hot

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.8Battery-faulty

The battery is faulty

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8418512.9Battery-incompatible

The battery is incompatible

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410608.0body-movement

Too much body movement

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410608.1cuff-too-loose

The cuff was too loose

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410608.2irregular-pulse

An irregular pulse was detected

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410608.3pulse-over-range-limit

The pulse was too high for sensor to work

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410608.4pulse-under-range-limit

The pulse was too low for sensor to work

http://terminology.hl7.org/CodeSystem/ASN1ToHL7  8410608.5improper-body-position

The cuff was incorrectly placed with respect to the heart