Personal Health Device Implementation Guide, published by HL7 International / Devices. This guide is not an authorized publication; it is the continuous build for version 3.0.0-ballot built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/phd/ and changes regularly. See the Directory of published versions
| Official URL: http://hl7.org/fhir/uv/phd/Requirements/fromNarrative | Version: 3.0.0-ballot | ||||
| Standards status: Trial-use | Maturity Level: 2 * | Computable Name: FromNarrative | |||
| Other Identifiers: OID:2.16.840.1.113883.4.642.40.73.36.1 | |||||
Conformance statements found throughout the narrative of the IG consolidated into this computable resource for traceability purposes
Language: en
| 1 | SHOULD | The PHG SHOULD specify a sufficiently selective conditional-create criterion so that at most one matching resource is found on the server. |
| 2 | SHALL | When an equivalent UCUM code is known, this UCUM code SHALL be used. |
| 3 | SHALL | An implementation SHALL use the MDC code if it does not know the UCUM code for the given unit. |
| 4 | SHALL | The converted effective[x] SHALL retain fractional-second precision consistent with the PHD's tick resolution, so that time differences between observations are preserved. |
| 5 | SHOULD | A derived RR interval Observation SHOULD reference the source ECG waveform Observation using Observation.derivedFrom. |
| 6 | SHALL | The MDC code SHALL be present in a coding element. |
| 7 | SHOULD | Because PHDs often have unreliably set clocks, a PHD that generates measurements with timestamps SHOULD provide a means for the PHG to determine the PHD's current time and clock status information, such as its synchronization state. |
| 8 | SHOULD | This measurement SHOULD contain a derived-from reference pointing to the measurement containing the glucose concentration. |
| 9 | SHALL | For a PHD, Device.conformsTo SHALL contain at least one MDC specialization entry. |
| 10 | SHOULD | PHDs SHOULD use the unchanged timestamp of a generated observation as received from the PHD. |
| 11 | SHOULD | PHGs compliant with this IG SHOULD implement this identifier as specified to ensure this scheme works across all compliant gateways. |
| 12 | SHOULD | For device settings known to the PHG, Observation.subject SHOULD point to the PHD. |
| 13 | SHOULD-NOT | For Coincident Timestamp Observations, Observation.performer SHOULD NOT be provided. |
| 14 | SHOULD | Observations whose timestamps are neither from the PHD's current timeline nor from a timeline synchronized with an external time source SHOULD be discarded, since there is no way to give them a correct timestamp. |
| 15 | SHALL | When the PHG modifies an Observation's timestamp as received from the PHD, it SHALL also generate a Coincident Timestamp Observation that records how the Observation.effective[x] element is generated. |
| 16 | SHOULD | When uploading to a FHIR server, the gateway SHOULD report the group observation with a hasMember reference to all member Observations. |
| 17 | SHOULD | In FHIR, the session Observation SHOULD have hasMember references to all observations that are part of the session. |
| 18 | SHALL | PHGs SHALL include a category element with a fixed coding as specified by the profile. |
| 19 | SHOULD | Vital-sign Observations SHOULD include a category element as defined by the Vital Signs profile. |
| 20 | SHALL | When encoding numeric Observation quantities in FHIR, the PHG SHALL follow the precision indicated by the PHD's IEEE 11073 FLOAT or SFLOAT encoding. |
| 21 | SHOULD | However, if the PHD does not report a System-identifier, which may happen if one needs to map non-certified and/or proprietary devices, a transport address SHOULD be reported. |
| 22 | SHALL | When a PHG encounters such a device it SHALL provide a transport address as a means of uniquely identifying the PHD. |
| 23 | SHALL | The Accuracy attribute SHALL be reported if the PHD provides it. |
| 24 | SHALL | Measurement-Confidence-95 SHALL be reported if the PHD provides it. |
| 25 | SHALL | There SHALL be one coding element using the MDC coding system. |
| 26 | SHALL | If the measurement is a vital sign, an additional coding element SHALL use a LOINC vital-sign code. |
| 27 | SHALL | The PHG SHALL use a Patient.identifier supplied by the server when one is provided. |
| 28 | SHALL | If the server does not provide an identifier, the PHG SHALL generate an implementation-dependent identifier whose value is unique within its assigning system. |
| 29 | SHOULD | The type for a PHG-generated identifier SHOULD use a suitable code from the HL7 v2-0203 coding system. |
| 30 | SHOULD | PHG implementers that generate device alerts SHOULD follow the two alert-generation pathways — threshold/limit alerts on numeric observations and device/sensor status-bit alerts — described in the PhdDeviceAlert profile. |
| 31 | SHALL | Uploads generated by PHGs compliant to this IG SHALL use the mapping defined by this implementation guide when transforming PHD data to FHIR resources. |
| 32 | SHOULD | The PHG SHOULD implement a duplication detection mechanism and filter out any observations that have already been uploaded. |