Personal Health Device Implementation Guide, published by HL7 International / Devices. This guide is not an authorized publication; it is the continuous build for version 3.0.0-ballot built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/phd/ and changes regularly. See the Directory of published versions

Example DeviceAlert: Phd Device Alert for a displaced sensor

Page standards status: Informative

Pulse Oximeter Sensor-Displaced Alert Example

Purpose

This example demonstrates a technical PhdDeviceAlert generated from a device/sensor status bit rather than from a measured value crossing a threshold. The pulse oximeter reported a device-status observation in which the sensor-displaced bit was set, and the Personal Health Gateway (PHG) raised a corresponding alert.

Key Features

1. Technical alert type

type = technical — the condition concerns the device/sensor, not a physiological value.

2. Condition code from the ASN1ToHL7 code system

code uses the device's own status code directly — 150604.2 sensor-displaced from the ASN1ToHL7 code system. The profile binds code to a value set that accepts either an IEEE 11073-10101 MDC partition-3 condition code or an ASN1ToHL7 code, so no MDC "umbrella" code is needed for a device/sensor status condition.

3. Derived from a bits-enumeration observation

derivedFrom.observation references bits-observation, a PhdBitsEnumerationObservation, and derivedFrom.component points at the exact ASN.1 bit (150604.2 sensor-displaced) that triggered the alert. There is no limit, because no threshold was crossed.

4. Signal

The alert is annunciated (signal.activationState = on, signal.presence = on).

How the sensor triggers this alert

Unlike a limit alarm (value out of range), a sensor-displaced condition is reported by the device as a status bit in its device-status object. The PHG reads that bit from the measurement report, surfaces it on the PhdBitsEnumerationObservation, and generates this DeviceAlert referencing that observation and the specific bit.