Personal Health Device Implementation Guide, published by HL7 International / Devices. This guide is not an authorized publication; it is the continuous build for version 3.0.0-ballot built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/phd/ and changes regularly. See the Directory of published versions
| Page standards status: Informative |
This example demonstrates a technical PhdDeviceAlert generated from a device/sensor
status bit rather than from a measured value crossing a threshold. The pulse oximeter
reported a device-status observation in which the sensor-displaced bit was set, and the
Personal Health Gateway (PHG) raised a corresponding alert.
type = technical — the condition concerns the device/sensor, not a physiological value.
code uses the device's own status code directly — 150604.2 sensor-displaced from the
ASN1ToHL7 code system. The profile binds code to a value set that accepts either an
IEEE 11073-10101 MDC partition-3 condition code or an ASN1ToHL7 code, so no MDC "umbrella"
code is needed for a device/sensor status condition.
derivedFrom.observation references bits-observation,
a PhdBitsEnumerationObservation, and derivedFrom.component points at the exact ASN.1 bit
(150604.2 sensor-displaced) that triggered the alert. There is no limit, because no
threshold was crossed.
The alert is annunciated (signal.activationState = on, signal.presence = on).
Unlike a limit alarm (value out of range), a sensor-displaced condition is reported by the
device as a status bit in its device-status object. The PHG reads that bit from the
measurement report, surfaces it on the PhdBitsEnumerationObservation, and generates this
DeviceAlert referencing that observation and the specific bit.
Profile: PhdDeviceAlert
status: In Progress
type: Technical
code: Pulse oximeter sensor displaced
subject: Sisansarah Lorianthah Piggy (no stated gender), DoB Unknown ( Medical record number: sisansarahId)
occurrence: 2018-11-11 19:07:48-0500 --> (ongoing)
| Observation | Component |
| Observation MDC_PULS_OXIM_DEV_STATUS | ASN1 bits to HL7 enumerated values: 150604.2 (sensor-displaced) |
label: Sensor displaced
| ActivationState | Presence | Manifestation | Type |
| On | On | Auditory | Sensor-displaced indication |