Personal Health Device Implementation Guide, published by HL7 International / Health Care Devices. This guide is not an authorized publication; it is the continuous build for version 1.1.0 built by the FHIR (HL7® FHIR® Standard) CI Build. This version is based on the current content of https://github.com/HL7/phd/ and changes regularly. See the Directory of published versions
Official URL: http://hl7.org/fhir/uv/phd/ValueSet/ASN1measurement | Version: 1.1.0 | |||
Active as of 2018-09-17 | Computable Name: ASN1measurement |
ValueSet for the ANS1ToHL7 codes that are derived from enumeration measurements.
References
Generated Narrative: ValueSet ASN1measurement
http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7
where source = measurement
Generated Narrative: ValueSet
Expansion based on codesystem ASN1 bits to HL7 enumerated values v1.1.0 (CodeSystem)
This value set contains 104 concepts
Level | Code | System | Display | Definition |
1 | 67846.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | lim-alert-off | Indicates both the high and the low limit alerts are disabled |
1 | 67846.1 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | lim-low-off | Indicates the low limit alert is disabled |
1 | 67846.2 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | lim-high-off | Indicates the high limit alert is disabled |
1 | 150604.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-disconnected | Indicates the device is not connected with the sensor |
1 | 150604.1 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-malfunction | Indicates the sensor has malfunctioned |
1 | 150604.2 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-displaced | Indicates the sensor is not correctly placed on the user |
1 | 150604.3 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-unsupported | Indicates the sensor connected to the device is unsupported |
1 | 150604.4 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-off | Indicates the sensor is not connected to the user |
1 | 150604.5 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-interference | Indicates the sensor is experiencing interference |
1 | 150604.6 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | signal-searching | Indicates the sensor is searching for a signal |
1 | 150604.7 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | signal-pulse-questionable | Indicates a questionable pulse has been detected |
1 | 150604.8 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | signal-non-pulsatile | Indicates a non pulsatile signal has been detected |
1 | 150604.9 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | signal-erratic | Indicates a signal has been discovered, but it is erratic |
1 | 150604.10 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | signal-low-perfusion | Indicates the signal is experiencing low perfusion |
1 | 150604.11 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | signal-poor | Indicates the signal has been discovered, but it is poor |
1 | 150604.12 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | signal-inadequate | Indicates the signal has been discovered, but it is inadequate |
1 | 150604.13 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | signal-processing-irregularity | Indicates the signal has been discovered, but is experiencing processing irregularity |
1 | 150604.14 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-equipment-malfunction | Indicates a device malfunction has occurred |
1 | 150604.15 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-extended-update | Indicates an extended display update is in progress on device |
1 | 150605.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | pulse-qual-nominal | Indicates no abnormalities have been detected by the sensor |
1 | 150605.1 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | pulse-qual-marginal | Indicates the pulse profusion or quality is marginal |
1 | 150605.2 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | pulse-qual-minimal | Indicates the pulse profusion or quality is minimal |
1 | 150605.3 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | pulse-qual-unacceptable | Indicates the pulse profusion or quality is unacceptable |
1 | 8410584.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | leadwire-loss | Loss of unspecified lead wire or electrode connection |
1 | 8410584.1 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | leadsignal-loss | Loss of unspecified lead signal |
1 | 8410584.2 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | leadwire-loss-first-lead | Loss of 1st lead wire or electrode connection |
1 | 8410584.3 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | leadsignal-loss-first-lead | Loss of 1st lead signal |
1 | 8410584.4 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | leadwire-loss-second-lead | Loss of 2nd lead wire or electrode connection |
1 | 8410584.5 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | leadsignal-loss-second-lead | Loss of 2nd lead signal |
1 | 8410584.6 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | leadwire-loss-third-lead | Loss of 3rd lead wire or electrode connection |
1 | 8410584.7 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | leadsignal-loss-third-lead | Loss of 3rd lead signal |
1 | 8417752.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-battery-low | The device battery is low |
1 | 8417752.1 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-malfunction | The sensor has malfunctioned |
1 | 8417752.2 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-sample-size-insufficient | The sensor's sample size is insufficient |
1 | 8417752.3 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-strip-insertion | The test strip was inserted incorrectly |
1 | 8417752.4 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-strip-type-incorrect | The test strip type is incompatable with the sensor |
1 | 8417752.5 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-result-too-high | The value reported by the sensor is too high |
1 | 8417752.6 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-result-too-low | The value reported by the sensor is too low |
1 | 8417752.7 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-temp-too-high | The temperature is too high |
1 | 8417752.8 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-temp-too-low | The temperature is too low |
1 | 8417752.9 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-read-interrupt | The read process was interrupted |
1 | 8417752.10 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-gen-fault | General Device Fault |
1 | 8417752.11 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-temp-out-of-range | The temperature is out of range |
1 | 8417909.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-device-battery-low | The INR device battery is low |
1 | 8417909.1 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-malfunction | The INR device sensor has malfunctioned |
1 | 8417909.2 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-sample-size-insufficient | The sample size was insufficient |
1 | 8417909.3 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-strip-insertion | The test script was inserted incorrectly |
1 | 8417909.4 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-strip-type-incorrect | The test strip used is not compatible with the INR |
1 | 8417909.5 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-result-too-high | The test result is too high |
1 | 8417909.6 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-result-too-low | The test result is too low |
1 | 8417909.7 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-temp-too-high | The temperature is too high |
1 | 8417909.8 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-temp-too-low | The temperature is too low |
1 | 8417909.9 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-read-interrupt | The read process was interrupted |
1 | 8417909.10 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-device-gen-fault | General device fault |
1 | 8417909.11 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | inr-sensor-calibration-due | The INR sensor calibration is due |
1 | 8408608.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-undetermined | Undetermined device status event |
1 | 8408608.1 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-reset | Device reset |
1 | 8408608.5 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-error | Device error |
1 | 8408608.6 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-error-mechanical | Mechanical error |
1 | 8408608.7 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-error-electronic | Electronic error |
1 | 8408608.8 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-error-software | Software error |
1 | 8408608.9 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-error-battery | Battery error |
1 | 8408608.15 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-service | Undetermined device status event, service needed |
1 | 8408608.16 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-service-time-sync-required | Time sync service required |
1 | 8408608.17 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-service-calibration-required | Calibration required |
1 | 8408608.18 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-service-replenishment-required | Replenishment required |
1 | 8408608.25 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-battery-low | Battery power is too low |
1 | 8408608.26 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-battery-depleted | Battery is depleted |
1 | 8408608.27 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-battery-replaced | Battery has been replaced |
1 | 8408608.28 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-status-battery-interrupted | Battery was interrupted |
1 | 8418060.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-session-stopped | Session stopped |
1 | 8418060.2 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-type-incorrect | Sensor type incorrect for device |
1 | 8418060.3 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-malfunction | The sensor malfunctioned |
1 | 8418060.4 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | device-specific-alert | Device specific alert |
1 | 8418060.7 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-calibration-not-allowed | Calibrating the sensor is not allowed |
1 | 8418060.8 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-calibration-recommended | Calibrating the sensor is recommended |
1 | 8418060.9 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-calibration-required | Calibrating the sensor is required |
1 | 8418060.10 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-temp-too-high | Temperature is too high for a valid result |
1 | 8418060.11 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-temp-too-low | Temperature is too low for a valid result |
1 | 8418060.12 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-result-below-patient-low | Sensor result below the Patient Low level |
1 | 8418060.13 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-result-above-patient-high | Sensor result above the Patient High level |
1 | 8418060.14 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-low-hypo | Sensor result below level |
1 | 8418060.15 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-high-hyper | Sensor result above level |
1 | 8418060.16 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-rate-decrease-exceeded | The sensor rate of decrease has been exceeded |
1 | 8418060.17 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-rate-increase-exceeded | The sensor rate of increase has been exceeded |
1 | 8418060.18 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-result-too-low | Sensor result is lower than the device can process |
1 | 8418060.19 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-result-too-high | Sensor result is higher than the device can process |
1 | 8418060.20 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | sensor-com-out-of-range | The sensor communication is out of range |
1 | 8418512.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-status-Undetermined | the status of the battery is unknown |
1 | 8418512.1 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-absent | Indicates if the battery is absent |
1 | 8418512.2 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-active | Indicates if the battery is active in the system |
1 | 8418512.3 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-charging | Indicates if the battery is charging |
1 | 8418512.4 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-fullyCharged | Indicates if the battery is fully charged |
1 | 8418512.5 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-disposable | Indicates if the battery is disposable |
1 | 8418512.6 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-rechargeable | Indicates if battery is rechargable |
1 | 8418512.8 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-faulty | Indicates if battery is faulty |
2 | 8418512.7 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-overTemperature | Indicates if the battery is too hot |
1 | 8418512.9 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | Battery-incompatible | Indicates if battery is incompatible |
1 | 8410608.0 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | body-movement | Indicates too much body movement |
1 | 8410608.1 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | cuff-too-loose | Indicates the cuff was too loose |
1 | 8410608.2 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | irregular-pulse | Indicates an irregular pulse was detected |
1 | 8410608.3 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | pulse-over-range-limit | Indicates the pulse was too high for sensor to work |
1 | 8410608.4 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | pulse-under-range-limit | Indicates the pulse was too low for sensor to work |
1 | 8410608.5 | http://hl7.org/fhir/uv/phd/CodeSystem/ASN1ToHL7 | improper-body-position | Indicates the cuff was incorrectly placed with respect to the heart |
Explanation of the columns that may appear on this page:
Level | A few code lists that FHIR defines are hierarchical - each code is assigned a level. In this scheme, some codes are under other codes, and imply that the code they are under also applies |
System | The source of the definition of the code (when the value set draws in codes defined elsewhere) |
Code | The code (used as the code in the resource instance) |
Display | The display (used in the display element of a Coding). If there is no display, implementers should not simply display the code, but map the concept into their application |
Definition | An explanation of the meaning of the concept |
Comments | Additional notes about how to use the code |